Installation and commissioning
PlasmaQuant 9100 (Elite)
36
}
Connect the pump hose for the purging solution to the lower intake con-
nection (1a) of the purging vessel. Place the pump hose over the hose
block from above and clamp it between two stoppers. Connect the intake
hose for purging solution to the other end of the hose (1b). Immerse the
intake hose in purging solution.
}
Connect the pump hose for waste to the top outlet connection (2a) of the
purging vessel. Place the pump hose over the hose block from below and
clamp it between two stoppers. Connect the waste hose to the other end
of the hose (2b). Insert the waste hose in the waste bottle.
NOTICE! Note the pump direction! The pump moves in a clockwise di-
rection.
}
Fasten the clamping bracket over the pump hoses via the clamping lever.
}
Insert the cannula into the holder on the sampler head.
– Move the holder along the Z-axis (down and up) and check whether
the cannula fits through the guide at the lower end of the head.
– Fasten the cannula on the holder with the nut (arrow in figure on the
left).
}
Initially guide the sample hose in an arc through the eyelet on the can-
nula holder (1).
}
Thread the hose from the left side through the eyelet (2) on the lower
end of the head.
}
Place the hose to the rear in the eyelets on the rear of the sampler arm.
}
Connect the hose with the sample hose of the basic device.
}
Check the DIP switch (1). Set switch 5 to "ON"; all other switches remain
at their initial positions.
}
Connect a USB cable to the "Host" connection and to the "Autosampler"
connection on the basic device.
}
Connect the power supply cable to the power connection (3) and to a
grounded power outlet.
}
Set the pump speed on the sampler so the liquid level remains constant
and not too much purging liquid overflows.