PlasmaQuant 9100 (Elite)
Installation and commissioning
31
}
Attach the torch to the carriage on the adjusting rail in the sample cham-
ber and screw it on.
NOTICE! Tighten the screws firmly so that the gas supply is leak-proof.
}
Fit together the spherical joint connection of the torch and the spray
chamber and secure the connection with the fork clamp.
}
Push the torch upwards on the adjusting rail until it engages with the
height adjustment.
}
Adjust the torch with the manual height adjustment (2) so that the tip of
the injector in the torch (1) is situated approx. 1 – 2 mm below the low-
ermost winding of the induction coil.
}
The outer tube must be removed to adjust the ceramic torch (HF kit).
– Adjust the torch with the manual height adjustment (2) so that the
outer edge of the inner tube is situated approx. 1 – 2 mm below the
lowermost winding of the induction coil.
– Reinsert the outer tube after adjustment.
}
Cut the sample and waste hoses to the required length, ensuring they are
long enough that the torch with the nebulizer unit can still move freely
on the adjusting rail. Slightly chamfer the ends of the hoses.
}
Connect the sample hose of the nebulizer to the pump hose with the
black stoppers and the waste hose to the pump hose with the red stop-
pers. Push the ends of the sample and waste hoses several millimeters
into the pump hoses.
To improve grip on the hoses, use a piece of fine sanding paper as a grip.