TOPwave
Operating
31
MAG
T.
Max. permissible magnetron temperature. If this temperature is exceeded
the digestion is aborted.
Permitted values: 0 -120°C
Typical values: 80 – 90°C
Values <20°C are used only for the venting programs.
Push [
ADD
]
to accept the application. Select [
SAVE
+
EXIT
], if you modified the applica-
tion.
The Main menu appears. The application is activated and can be used for the next di-
gestion.
Note:
Only user with
MASTER
login can modify application. Standard applications cannot be
modified.
Select an application in the library and push [
MODIFY
].
Modify the parameters (see above).
With [
SAVE
+
E
XIT
]
confirm the changes and return into the Main menu. The changed
application is used for the next digestion.
Of the 200 applications in the selective list 30 are pre-installed and 25 are currently al-
located.
Program number
Description
1
Calcite, Dolomite
2
Cleaning, Vessel conditioning
3
Coal, DIN 22022-1
4
Complexation H3BO3
5
Evaporation
6
Feed
7
Feldspar
8
Filter, glass fiber
9
Fly Ash, DIN EN 14385
10
Food (no/low fat)
11
Food (High fat)
12
Glass, Quartz
13
Limestone
14
Oil, Lubricant
15
Plastic, PE, ROHs
16
Plastic, Waste DIN 22022-1
17
Soil ISO 11466, EN 13346
18
Soil EPA 3051
19
Soil, Sediment EPA 3052
20
Tissue, Blood
21
Vegetables, Leaves
22
Vegetable Oil
Modifying application in
the library
List of pre-installed stan-
dard programs
1 – 25