TOPwave
Installation and Setup
21
5
Installation and Setup
5.1
Unpacking
Open the shipping packaging and carefully remove it. Please save all packaging in order
to be able to return the unit to the manufacturer in its original packaging, should service
be required.
Two persons should remove the microwave system from its packaging and place it on a
lab bench. Also remove all other components and accessories and inspect the delivery to
make sure it is complete.
Inspect the system for shipping damage such as cracks, scratches, dents, etc.
Should any components be missing or damages be detected, contact your local, authori-
zed dealer or Analytik Jena directly.
WARNING
Visible damages!
For safety reasons, never operate the unit if it exhibits visible signs of damage. Contact
your local, authorized dealer or Analytik Jena to obtain the required service.
5.2
Installing TOPwave
WARNING
The unit employs high voltage and generates microwave radiation! The TOPwave must
always be carefully connected to a grounded conductor. Provided the wall outlet used is
equipped with a ground connection, the supplied power cord will ensure proper connec-
tion. If it is necessary to use an extension cord to connect the system to the power
supply, only a three-wire cord with a ground connection may be employed.
The unit should not be set up near any devices which are sensitive to electromagnetic
radiation. If necessary, a warning sign or label indicating “Caution, Microwave Emissions”
should be applied or attached to the unit. Should there be any suspicion or indication of
increased levels of microwave radiation, please immediately inform the Service Depart-
ment at Analytik Jena.
Repair and servicing of the equipment may only be performed by trained personnel.
Position the system on the intended work surface. Make sure the vent opening on the
left side of the unit is not covered (at least 15cm (5.9inch) clearance). A clearance of at
least 50 cm (19.7 inch) that opens towards the front of the unit must be available above
the oven. The work surface must be able to support the weight of the oven (approx. 80
kg; 176 lb) and must have a surface area of approx. 90 x 80x 90 cm (35,4 x 31,5 x 35,4
inch; W x D x H).
It is possible to feed or suck off acid fumes directly from the collecting vessel or pressure
vessels to an air extraction system via the PFA hose, which can be connected through
the lid from above, (e.g. by means of a membrane or water-jet suction pump with an
Positioning
Exhaust