AOX determination
multi X 2500
72
6.3.3
Performing measurements
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear appropriate
protective clothing when working at the sulfuric acid container. Observe all
specifications in the safety data sheet!
1.
Fill fresh sulfuric acid into the sulfuric acid container with outlet on the chlorine
module every day (
→
"Changing the sulfuric acid and cleaning the sulfuric acid
container with outlet", p. 101).
2.
Preparing the measuring cell:
Fill the measuring cell with fresh electrolyte solution every day.
Measuring cell "sensitive":
15 to 20 ml
Insert the magnetic stirring rod into the measuring cell. Attach the cover and
tighten it using the three knurled head screw.
Insert the combined electrode and connect it electrically. Electrically connect
the measuring cell.
Connect the combined electrode to the hose of the sulfuric acid container.
Connect the olive-tip suction adapter to the exhaust air hose.
NOTICE! Never connect the measuring cell to the intake pump via the olive-tip
suction adapter and the hose. Otherwise, electrolyte will be drawn into the pump
unit.
3.
The analyzer and the chlorine module are switched on.
The devices are booted. The status LEDs on the front panel lights up in green
after approximately 30 seconds.
4.
Open the gas supply and set the primary pressure to 4 to 6 bar.
5.
Switch on the PC.
6.
Start the control and analysis software multiWin and log in with user name and
password.
7.
Initialize the analysis system by clicking on [
INITIALIZE ANALYZER
].
The initialization and automatic detection of all connected components will be
executed.
In the window D
EVICE STATUS
any components not yet ready for operation will be
shown in red. The time it takes to preheat the furnace to 950 °C is about 30 to
40 min. During the warm-up phase it is not possible to start a measurement.
If the analyzer is still not ready for measurements after about 40 min., carry out a
fault analysis in accordance with the notes in chapter "Fault removal" on page 106.
8.
Carry out an end point routine.
9.
If necessary, test the measuring cell using a direct method.
10.
Providing the samples:
Load the AOX/solids rack of the autosampler autoX 112 with samples.