multi X 2500
Fault removal
111
272
Auto-injector – syringe: incorrect size
Cause
Remedy
Dosing volume and syringe size differ
Adjust dosing volume and syringe size
273
Auto-injector – syringe: not filled properly
Cause
Remedy
Syringe not drawn all the way to the stop
Draw syringe completely
Insert syringe
274
Auto-injector: no connection
Cause
Remedy
Auto-injector coupling was not found
Check connection
Acknowledge message
Initialize
300
Temperature controller: no connection
Cause
Remedy
Communication error
Acknowledge message
Initialize
304
Temperature controller: communication error
Cause
Remedy
Temperature cannot be set
Acknowledge message
Initialize
Communication error
Acknowledge message
Initialize
11.2
Equipment faults and analytical problems
11.2.1
General notes
Other problems not detected by the system monitoring can also occur. Starting a
measurement is possible. Such errors are usually detected on the basis of measuring
results which seem to be implausible (analytical problems) or they are clearly indicated
by the equipment technology.
If the suggested solutions are not successful, inform the customer service department
at Analytik Jena.
11.2.2
Equipment fault
furnace does not heat
Cause
Solution
Thermocouple connector not connected
Temperature set incorrectly in the software
No method loaded
Malfunction in power supply
Malfunction in the internal electronics
Connect thermocouple connector (4 in Fig. 5, p. 20)
Check temperature configuration in the method
Load method
Switch on the device,
check the internal fuse
check the connection b/w basic device and PC
Report to customer service