InnuPure C16
touch
Function and setup
19
Fig. 3
Device view with open front door
1 Sample tray counter-holder
2 Drawer
3 Protection cover of the combined magnet/heating
element
4 Cones tip attachment
During operation, the front door of the InnuPure C16
touch
is closed. The device is
equipped with an LED for status control. Depending on the operating status, the LED
illuminates green or red.
All pre-manufactured reagent packs, the filter tips and the elution tubes required for
the extraction process are placed onto the sample tray. Depending on the number of
samples, reagent plates (deep well plates) with eight sample positions or reagent
strips for individual samples will be used. Thus, within a single extraction protocol, in-
dividual samples in reagent strips or up to 16 samples in two reagent plates can be
processed. The optionally available adapter is required to accept the reagent strips.
Reagent plates are placed directly onto the sample tray. For populating the sample
tray with plates and strips, the tray must be placed into the priming station supplied.
1 Holding-down clamp
2 Sample block for reagent
plates or adapter for rea-
gent strips
3 Tip block
4 Holder for tips and elution
tubes
Fig. 4
Sample tray