1. Technical specifications
Copyright ©
ŘŖŗŜ
by Endecotts Ltd.
57
1. Technical specifications
x
Check the degree of soiling of the vacuum cleaner filter regularly and change the filter
where applicable.
x
The vacuum or differential pressure generated by the vacuum cleaner may otherwise be
too low for sieving.
x
Maintenance
This device is essentially maintenance-free if cleaned regularly.
E11 MOTOR
FAULTY/BLOCKED
Service required
E83
VACUUM TOO LOW
Check whether
•
the suction apparatus is connected;
•
the suction apparatus is generating
sufficient vacuum;
•
The collecting receptacle in the suction
apparatus is full;
•
the sieve cover is attached.
E84
DROP IN VACUUM
Check whether
•
the sieve cover is attached;
•
the suction hose has been connected;
•
the vacuum cleaner bag is full.
QAQC LAB 589 Rappahannock Drive White Stone Va 22578 TEL (804) 318-3686