System Maintenance and Troubleshooting
EM3 AEM Monitor User’s Guide
4-3
Electrical Inspection of the Bipolar Current Indicator
1. Use a Bipolar Adapter to connect the AEM Monitor to a compatible electrosurgical
generator with a continuously adjustable output.
2. Connect a noninductive load resistor (between 50 ohms and 200 ohms) and a reference
ammeter in series to the Bipolar Instrument receptacle of the AEM Monitor.
The reference ammeter should have a true RMS response, a bandwidth of at least 10 MHz,
and an accuracy of 1%.
3. Agreement between the End Point Monitor and the reference meter should be within 20%
of display with currents between 600 mA and
1000 mA.
Mechanical Inspection
Visually inspect the following items on the AEM Monitoring System and the End Point Monitor. If
any of these items appear damaged during the visual inspection, contact the Encision Service
Center.
Insulation of wiring and cables
Connectors and cables are fully seated
All hardware is securely fastened
Monopolar Adapter cable and plug
Monopolar Adapter receptacle and pins
Instrument receptacles, sockets, and pins on front panel
Front panel keypad buttons and LED indicators
Power switch
Power entry module
Cleaning the AEM Monitor
WARNING
Electric Shock Hazard. Always unplug the AEM Monitor before cleaning.
Clean monitor when needed.
1. Turn off the AEM Monitor.
2. Disconnect all accessories.
3. Follow the procedures approved by your institution or use a validated infection control
procedure.
4. Use a mild cleaning solution (100:1 water to mild detergent, by mass) or disinfectant and a
damp cloth to thoroughly wipe all outside surfaces and the power cord.
CAUTION
Do not allow fluids to enter the chassis.
Do not clean the AEM Monitor with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the unit.
Do not steam sterilize the AEM Monitor.
Storing the AEM Monitor
The AEM Monitor should be stored within these parameters:
Temperature: -13 to 158° F (-25 to 70° C)
Humidity: 15% to 95% relative, non-condensing
If you store the AEM Monitor at a temperature that is outside its normal operating range of 59 to 104° F (15
to 40° C), allow one hour for the monitor to reach room temperature before use.