Users’ Manual
EAN-90M Tiltmeter
Page
|
7
3.4 Connecting sensor to EDAS-10 data acquisition system
Model EAN-90M tilt meter requires external power source of ± 12V DC (regulated) when connected to
EDAS-10 data acquisition system. For connecting tilt meter, a multiplexer model EAM-616 is required. 16
tilt meters can be connected to each multiplexer model EAM-616.
Maximum input of EDAS-10 is ± 2.5 V; whereas output of tilt meter is ± 4 V at 15
0
proportional to Sin of
angle. A high precision attenuator is therefore required as shown in diagram below:
Use EDAS-10 function P2 to read output of tilt sensor in terms of 0.5 volts. The multiplier and offset should
be kept as 1 and 0 respectively. To get tilt angle refer to typical gage factor and formula given in relevant
test certificate. A sample test certificate for uni-axial sensor can be referred to in section 3.5.
P S card
+12V
G
-12V
1
2
3
EAM 616 2C8
1
3
2
Green
Black
Red
-12V
+12V
0V
4
6
5
Datalogger
H
Blue
A AXIS
B AXIS
White
Brown
COM
Tilt sensor
Tilt measurement with Tilt sensor and DAS
S
W
-12V
G
4
6
5
COM
CH.
*
*
Required for biaxial sensors only.
R1 = R2 = R3 = R4 = 100k ±0.01%, 5ppm.
*
*
R1
R2
R3
R4
*
*
L
SE
Figure 5: Tiltmeter connected to DAS