Patient Compliance Timer
Operation count: record of 60 sets (min.), max with 999 mins.
Operation total time: max with 999 hrs.
Operation ambient:
Temperature range: 10°C ~ 35°C
Humidity range: 20 ~ 90%RH
Storage & transportation:
Temperature Range: 0°C ~ 70°C
Humidity Range: 20 ~ 90%RH
Timer:
5~90 minutes auto-shutoff or Constant
Mode
BURST
NORMAL
MRW
SD
BI-PULSE
Pulse Rate
Fixed 100Hz
2~150Hz
2~150Hz
2~150Hz
Channel 1
fixed at 4Hz;
Channel 2
fixed at 100 Hz
Pulse Width
50
~
300
µs
50
~
300
µs
50
~
300
µs
50
~
300
µs
50
~
300
µs
Cycle Time
0.5 Sec
Constant
1 Sec.
6 Sec.
2 Sec.
9.0
| TENS Treatment Functions Descriptions
8.0
| Specifications continued