57
58
Causes Solutions
l
Make sure the test strip vial has not been opened for
longer than three (3) months.
l
Make sure the meter code and test strip code match.
The
thermometer
icon ( ) appears
on the screen.
l
Place the meter and test strip in a temperature
environment of 50-104 . You should use it after about 20
minutes in the appropriate conditions.
l
If the thermometer icon is constantly showing, contact
Customer Service at 1-888-885-6677.
℉
The meter does
not work after
the test strip is
inserted.
l
Make sure the test strip is completely inserted into the test
strip port.
l
Make sure the batteries are inserted correctly, with the +
sign up and facing the right direction.
Technical Information and Warranty
Specifications
Product Name
Test Sample
Calibration
Sample size
Test Strip Type
Measuring Range
Measuring Time
Power
Battery Life
Memory Capacity
Data Transport
Medisign MM1100 BT Blood Glucose Monitoring System
Fresh capillary whole blood
Plasma-equivalent
0.5
㎕
Medisign MM1100 Blood Glucose Test Strip
20 - 600 mg/dL
5 Seconds
(2) 3.0 Volt CR2032 lithium batteries
Approximately1000 tests
300 test results
Wireless (Bluetooth 4.0), USB Data Transfer Cable