User manual motion sprint 600 med /600
Version 1.1
Page 15 of 21
Distance/ Distance Training
For this program, the parameters distance, speed and for SL inclination must be pre-selected. At any time during
the workout, the speed can be adjusted with
+/-
. The chosen distance will countdown and the intensity is
determined from among the 21 intensity levels. The workload is shown in km/h in the display. The time that is
needed to complete the distance can be used as an indicator of performance.
After completing the program, a display of the results is initiated.
Isokinetic/Power Training
This program is not available for the treadmill.
Watt sensitive/ not available with fitness program packages
Speed constant training programs are likely to be used in the field of training controlling and therapy, because
the intensity controlling can occur very accurate.
However, speed constant controlling can lead to overstress. To eliminate these negative aspects, a too high
stress gets avoided in the watt sensitive program with an intelligent program controlling. The overshoot of the
limited heart rate gets controlled. When it gets overshot for a longer time a speed reduction of 10% every 30
seconds follows.
The target speed/inclination will be reached via a ramp during the first 2 minutes. That way the trainee gets
used to the power and an adequate warm-up is assured.
Therapy/ not available with fitness program packages
Also in rehabilitation the total intensity of the training gets controlled with the intensity steering factors
amount as well as intensity, alongside the recovery progress. As a general rule intensity changes occur by
increasing the intensity. At the therapy training the intensity gets varied several times within the program.
After a two minute warm-up in ramp shape, 6 intervals of one minute each, with one minute active rest (50%
of the impact intensity) per interval follow.
Also the overshooting of the pre-entered heart rate gets controlled. When overshooting for a longer time a
reduction of the speed of 10% every 30 seconds fallows.
A Cool Down function over 3 minutes rounds up the training. Overall the program lasts 17 minutes.
4.13 Pulse controlled programs
Concerning the working of pulse controlled programs such as Pulse program/Pulse, Watt sensitive and Therapy,
please refer to the brochure “motion pulse manager”.
4.14 Interface PC
All ergometers are equipped standardly with an RS232 interface. This interface is not equipped with a galvanic
isolation or external power supply. By this the ergometer can be controlled via a PC, an ECG, spiroergometry
device, aso, if both devices are using the same interface protocol. Please note, that with medical use, the rules
for electric safety must be obeyed. Optionally, a RS232 interface with galvanic isolation can be ordered (when
the machine is being build), in order to limit the risk resulting from any malfunction of the connected device.