EMIT TB-6568 Скачать руководство пользователя страница 4

EMIT 

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 664-9980 • Fax (909) 627-7449 • Website: 

DescoEMIT.com

TB-6568

 Page 4 of 6

© 2009 DESCO INDUSTRIES, INC.

Employee Owned

HOLDING THE LAST READING

Press the HOLD button to freeze the reading from the 

object on the display and the analog output signal.  This 

feature allows the operator to move the Meter where it 

may be more easily read or saved for later reference.

made at the location where ESD sensitive items are 

to be ionized. Air ionizer heaters and air filters (if so 

equipped) should be left in their normal conditions during 

test.”

ANALOG OUTPUT

The analog output jack labeled “OUT” on the face of 

the Meter accepts a standard 2.5 mm monaural phone 

plug and is provided so the output of the Digital Static 

Field Meter may be connected to an oscilloscope, strip 

chart recorder, external meter or other device.  Use the 

included cord to achieve a connection between the Field 

Meter and alternate measuring instrument.  The voltage 

at this output is 1/1000th (±2 kV range) or 1/10,000 (±20 

kV range) of the measured voltage.

TAKING DISCHARGE TIME MEASUREMENTS

In order to verify that an ionizer is operating properly it is

also important that its ability to neutralize or discharge 

static electricity is measured.  The following procedure 

will measure an ionizer’s discharge time:

OPERATING THE CHARGER

The Charger has a momentary rocker-switch that powers

the unit.  Holding the switch forward / backward supplies

power to the output terminals.

POLARITY SELECTION

The top of the rocker switch is labeled “+”, and the 

bottom is labeled “-”.  To provide a POSITIVE voltage 

output, touch the plate located underneath the charger, 

and press the switch forward at the same time.  To 

provide a NEGATIVE voltage output, touch the plate 

located underneath the charger, and press the switch 

downward at the same time.

Note:

 For the Charger to work correctly, the operator 

and Field Meter must be properly grounded.  A ground 

path to the touch plate must exist.

IONIZER DISCHARGE TIME MEASUREMENTS

Use the Field Meter with the conductive plate in the

appropriate location for measurements.

POSITIVE DISCHARGE TIME MEASUREMENTS

To provide a POSITIVE voltage output, touch the plate

located underneath the Charger, and press the switch

forward at the same time.  Momentarily touch the 

Charger’s output terminal to the conductive plate 

attached to the Field Meter.  The meter reads 

approxi1.10 kV.  By using a stop watch or other 

timing device, determine the time needed for the 

voltages to decrease from +1.10 kV to +0.10 kV.  This is 

the positive discharge time.

Figure 7. Reading the Digital Static Field Meter while in 
the ±2 kV range

Hundreds
Tens
Ones

Figure 6. Reading the Digital Static Field Meter while in 
the ±20 kV range

Thousands
Hundreds
Tens

Per ESD TR53-2006 Compliance Verification of ESD

Protective Equipment and Materials Air Ionizer Test

Procedure Initial Test Setup “Measurements should be 

© 2010 DESCO INDUSTRIES INC.

Employee Owned

Содержание TB-6568

Страница 1: ...nsulators in the ESD protected area can adversely impact quality productivity and reliability When any object becomes electrostatically charged there is an electrostatic field associated with that charge If an ESDS ESD sensitive device is placed in that electrostatic field a voltage may be induced on the device If the device is then momentarily grounded a transfer of charge from the device occurs ...

Страница 2: ...nt operation B RANGE ZERO Button Press to select the measurement range Press and hold to zero the Meter C Analog Output Jack A low voltage replica of the measured voltage is provided at this output The voltage at this output is 1 1000th 2 kV range or 1 10 000 20 kV range of the measured voltage D POWER Button Press to turn the unit ON and OFF FRONT VIEW BACK VIEW 1 2 3 4 Figure 3 Charger features ...

Страница 3: ...rate measurements INSTALLING THE ISOLATED PLATE ASSEMBLY The Digital Static Field Meter s case has two slots along its sides The top slot is closest to the face of the instrument Slide down the tabs of the Conductive Plate plate into the top slot of the Meter s case as far as they go see Figure 4 BATTERY CHECK The battery should be replaced when BAT is indicated on the display Always replace the b...

Страница 4: ...h that powers the unit Holding the switch forward backward supplies power to the output terminals POLARITY SELECTION The top of the rocker switch is labeled and the bottom is labeled To provide a POSITIVE voltage output touch the plate located underneath the charger and press the switch forward at the same time To provide a NEGATIVE voltage output touch the plate located underneath the charger and...

Страница 5: ...isopropyl alcohol Then allow the instrument to air dry for several hours Specifications DIGITAL STATIC FIELD METER Measurement Range switch selectable Low Range 0 to 1 99 kV inch High Range 0 to 19 99 kV inch Measurement Accuracy Voltage Monitor Output 5 of reading 10 mV Voltage Display 5 of reading 2 counts Measurement Stability 10 counts Automatic Shutoff Unit will shut off after 20 minutes afte...

Страница 6: ... Return Material Authorization RMA and proper shipping instructions and address EMIT will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warran...

Отзывы: