2. Rocker Switch:

 Press and hold to select the polarity 

that will be provided at the Output Contact.

3. Touch Plate:

 Make contact with the touch plate while 

pressing down the rocker switch to provide voltage to the 

Output Contact.  The operator must be properly 

grounded during use.

4. Battery Compartment:

 Slide the cover down to open 

the 9V battery compartment.

Operation

TAKING OFFSET VOLTAGE (BALANCE) 
MEASUREMENTS

The Ionization Test Kit has been designed to match the

compact size and hand held convenience of the Digital

Static Field Meter.  Use the following procedure to verify 

the offset voltage (balance) of air ionization equipment.  

This quick and easy procedure will help determine if 

the piece of ionization equipment is working within the 

manufacturer’s specifications or user requirements.  

It 

is extremely important that ionizers be checked 
regularly for offset voltage (balance) and discharge 
times.  An ionizer operating in an out-of-balance 
state can place a charge on sensitive electronic 
components or assemblies.

Note: 

The 

50597

 Digital Static Field Meter is built in a 

conductive case.  The instrument senses the difference 

in potential between the case (and the person holding 

the case / ground connection) and the surface under 

test.  Ensure that the person using the instrument is 

grounded or that the rear panel ground snap connection 

is utilized to achieve accurate measurements.

INSTALLING THE ISOLATED PLATE ASSEMBLY

The Digital Static Field Meter’s case has two slots along 

its sides.  The top slot is closest to the face of the 

instrument.  Slide down the tabs of the Conductive Plate 

plate into the top slot of the Meter’s case as far as they 

go (see Figure 4).

BATTERY CHECK

The battery should be replaced when “BAT” is indicated 

on the display.  Always replace the battery with a 9V 

alkaline or equivalent battery in order to remain CE 

compliant.

ZERO THE METER

Turn the Meter on by pressing the POWER button.  

Press the RANGE / ZERO button to set the Meter to 

the 2 kV (3 decimal places) range.  Point the top of the 

Meter approximately 1 inch away from a grounded metal 

surface.  Use the red LED range guide.  The Meter is 

properly positioned when the projected red bullseyes 

are centered on top of each other.  Press and hold the 

RANGE / ZERO button until the Meter displays “.000”.

MAKING A MEASUREMENT

Locate the Test Kit in an ionized environment at the

appropriate distance from the device under test.  The 

static field displayed is the actual balance of the ionizer 

or voltage offset.  The display will indicate “1” or “-1” 

when the Meter is over-ranged.  Change the range of the 

unit if necessary. (see Figure 3).

Note:

 When testing pulsed ionizer systems, the voltage

displayed is constantly changing.  This pulse rate may 

be faster than the display update rate of the Field Meter,

therefore the displayed voltage is an average of the 

actual voltage.  The output of the Field Meter is useful in 

this situation for more accurate measurements.

Figure 4. Installing the 

50567

 Conductive Plate

12"

18"

Figure 5. Auditing ionization equipment with the Digital 
Static Field Meter and Conductive Plate (Ref: ANSI/ESD 
SP3.3)

Содержание TB-6568

Страница 1: ...nsulators in the ESD protected area can adversely impact quality productivity and reliability When any object becomes electrostatically charged there is an electrostatic field associated with that charge If an ESDS ESD sensitive device is placed in that electrostatic field a voltage may be induced on the device If the device is then momentarily grounded a transfer of charge from the device occurs ...

Страница 2: ...nt operation B RANGE ZERO Button Press to select the measurement range Press and hold to zero the Meter C Analog Output Jack A low voltage replica of the measured voltage is provided at this output The voltage at this output is 1 1000th 2 kV range or 1 10 000 20 kV range of the measured voltage D POWER Button Press to turn the unit ON and OFF FRONT VIEW BACK VIEW 1 2 3 4 Figure 3 Charger features ...

Страница 3: ...rate measurements INSTALLING THE ISOLATED PLATE ASSEMBLY The Digital Static Field Meter s case has two slots along its sides The top slot is closest to the face of the instrument Slide down the tabs of the Conductive Plate plate into the top slot of the Meter s case as far as they go see Figure 4 BATTERY CHECK The battery should be replaced when BAT is indicated on the display Always replace the b...

Страница 4: ...h that powers the unit Holding the switch forward backward supplies power to the output terminals POLARITY SELECTION The top of the rocker switch is labeled and the bottom is labeled To provide a POSITIVE voltage output touch the plate located underneath the charger and press the switch forward at the same time To provide a NEGATIVE voltage output touch the plate located underneath the charger and...

Страница 5: ...isopropyl alcohol Then allow the instrument to air dry for several hours Specifications DIGITAL STATIC FIELD METER Measurement Range switch selectable Low Range 0 to 1 99 kV inch High Range 0 to 19 99 kV inch Measurement Accuracy Voltage Monitor Output 5 of reading 10 mV Voltage Display 5 of reading 2 counts Measurement Stability 10 counts Automatic Shutoff Unit will shut off after 20 minutes afte...

Страница 6: ... Return Material Authorization RMA and proper shipping instructions and address EMIT will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warran...

Отзывы: