TB-6608
Page 4 of 5
© 2021 DESCO INDUSTRIES INC
Employee Owned
DESCO EUROPE
- 2A DUNHAMS LANE, LETCHWORTH, HERTFORDSHIRE, SG6 1BE, UK
Phone: +44 (0) 1462 672005 • E-mail:
Calibration
Frequency of recalibration should be based on the
critical nature of those ESD sensitive items handled and
the risk of failure for the ESD protective equipment and
materials. In general, EMIT recommends that calibration
be performed annually.
Use the EMIT 50424
Limit Comparator and EMIT 50784
5-Pound Electrode to perform periodic verification (once
every 6-12 months) of the SmartLog Pro® SE. The Limit
Comparator can be used to check the test limits of the
SmartLog Pro® SE without removing it from the turnstile
or factory floor.
See
Figure 7. EMIT 50424 Limit Comparator
Figure 8. EMIT 50784 5-Pound Electrode for Limit
Comparator
Specifications
SmartLog Pro® SE with Turnstile
SmartLog Pro® SE
Input Voltage and
Frequency
(External Adapter)
AC/DC Power Adapter
Power Input:
100-240VAC, 50/60 Hz
Power Output:
5VDC, 3.0A
Cable Length:
5 ft. (1.5 m)
Turnstile
Input Voltage and
Frequency
220VAC, 50/60 Hz
Operating Temperature 70°F to 85°F (21°C to 30°C)
for 1 gigohm test limit
41°F to 85°F (5°C to 30°C)
for all other test limits
Environmental
Requirements
Indoor use only at altitudes
less than 6500 ft. (2 km)
Maximum relative humidity
of 80% up to 85°F (30°C)
decreasing linearly to 50% @
85°F (30°C)
Maximum relative humidity of
50% at 1 gigohm setting
Dimensions (Crated)
47" x 31" x 54"
(119 cm x 78 cm x 137 cm)
Dimensions (Uncrated) See Figure 9
Weight (Crated)
257 lbs (116 kg)
Weight (Uncrated)
135 lbs (61 kg)
Test Accuracy
±20% for 1 gigohm footwear
test limit
±10% for all other test limits
Test Switch Voltage
5 VDC @ open circuit
Wrist Strap and
Footwear Test Voltage
30 VDC @ open circuit
Test current is limited by
resistors and varies on
the test range setting (100
kilohms - 1 gigohm)
Dual Independent Foot Plate
Dimensions
14.0" x 16.0" x 0.9"
(35.6 cm x 40.1 cm x 2.3 cm)
Weight
7.5 lbs (3.4 kg)