EMIT
- 3651 Walnut Avenue, Chino, CA 91710 • (909) 664-9980 • Website:
TB-6512
Page 3 of 5
© 2014 DESCO INDUSTRIES, INC.
Employee Owned
TESTING THE MAT CIRCUITS
1.
Connect the Calibration Unit’s GROUND test lead to a
ground point.
2. Insert a wrist cord into the satellite’s operator jack and
snap it to the parking station. This will enable the
audible alarm circuit needed for calbration verfication.
3. Connect the included stacking snap to the Calibration
Unit’s MAT test lead. Disconnect the satellite from
its mat connection and re-install it to the Calibration
Unit’s stacking snap.
NOTE: The stacking snap must be isolated from the
grounded mat.
Figure 5. Connecting the MAT test lead from the Monitor
Calibration Unit to the satellite
4. Rotate the Calibration Unit’s selector knob to the MAT
FAIL 10M position. The monitor’s corresponding red mat
LED should illuminate, and the alarm should sound.
5. Rotate the selector knob counter-clockwise to the MAT
PASS 10M position. The monitor’s corresponding green
mat LED should illuminate.
6. Perform the same procedure for the other satellite.
Calibrating the Jewel
®
Workstation Continuous
Mini Monitor
The Jewel
®
Workstation Continuous Mini Monitor is available
as the following item number:
TESTING THE OPERATOR CIRCUIT
1.
Connect the Calibration Unit’s GROUND test lead to a
ground point.
2. Insert the Calibration Unit’s OPERATOR test lead into
the Jewel
®
Mini Monitor’s OPERATOR jack.
Figure 6. Connecting the OPERATOR test lead from the
Monitor Calibration Unit to the Jewel
®
Workstation
Continuous Mini Monitor
3. Start the Calibration Unit’s selector knob at the full
counter-clockwise position. When testing the operator
circuit, test only the first four limits on the Calibration
Unit. Rotate the selector knob to the LOW FAIL position.
The red OPERATOR GROUND LED on the monitor
should illuminate, and the alarm should sound.
4. Rotate the selector knob clockwise to the LOW LIMIT
position of the green pass sector. The monitor’s green
OPERATOR GROUND LED should illuminate.
5.
Rotate the selector knob clockwise to the HIGH PASS
position at the end of the green pass sector. The
monitor’s green OPERATOR GROUND LED should
remain illuminated.
6. Rotate the selector knob clockwise to the HIGH FAIL
position. The monitor’s red OPERATOR GROUND LED
should illuminate, and the alarm should sound.
7. Disconnect the Monitor Calibration Unit from the monitor.
Brand
Item
Charleswater
99135