Instruction Manual Rev B
46
RS485: Pin out
Use RS485 to create a multi-drop network containing up to 32 devices.
Note: port #1is set to RS485
Note that the ports are not isolated. If controller is
used with a non-isolated external device, avoid
potential voltage that exceeds +/- 10V. To avoid
damaging the system, all non-isolating device ports
should relate to the same ground signal.
Use shielded, twisted pair cables.
Minimize the stub (drop) length leading from each device to the bus.
Ideally, the main cable should be run in and out of the network device.
Do not cross positive (A) and negative (B) signals. Positive terminals must be wired to positive and negative
terminals to negative.