Instruction Manual Rev B
40
Trouble Shooting
Problem
Possible Cause
Remedy
Control unit reads high opacity
1- Smoke
2- Alignment is out
3- Transceiver and/or retro
window is dirty
1- Correct process
2- Adjust alignment until centered on
target.
3- Clean windows.
Control unit reads 100%
Transceiver current loop to
the control unit is open
Check wiring for open from Terminal
0V & I6 on control unit to wire 9 & 8
at J-box location.
Alignment is good but control unit reads high
opacity.
1- Reference voltage TP-2
on signal processor 1667 is
lower than 9.3V
2- Main lamp out
1- Adjust lamp drive PT-2 on the
power modulator 1668 until
Reference voltage TP-2 on signal
processor 1667 is 10.0 Volts +/- .3v.
2- Replace main lamp assembly
Control unit reads High, lens clean, alignment
is good.
Dirt built up in flanges
Swing open transceiver & Retro.
Clean flanges with push rod.