3. For additional troubleshooting information, select the desired message and press
ENTER
.
Troubleshooting when a Fault message is showing
.
4. To return to the main display, press
MENU
and then
EXIT
.
9.3
Troubleshooting when a Fault message is
showing
Fault message
Explanation
Section
Main Board CPU Error
Main board software is corrupted.
Main Board CPU, Main Board Factory
Data, and Main Board User Data errors
Main Board Factory Data
Main board factory eeprom data is
corrupted.
Main Board CPU, Main Board Factory
Data, and Main Board User Data errors
Main Board User Data
Main board user eeprom data is
corrupted.
Main Board CPU, Main Board Factory
Data, and Main Board User Data errors
Sensor Hardware Error
Missing or bad hardware component.
Sensor Board Unknown
Transmitter does not recognize sensor
board.
Sensor HW-SW Mismatch
Sensor board hardware and software
do not match.
Sensor Incompatible
Sensor board software is not
supported by main board software.
Sensor Not Communicating
Sensor board is not communicating
with main board.
Sensor CPU Error
Sensor board software is corrupted.
Sensor RTD Open
Temperature measuring circuit is
open.
S1 Not Detected
No sensor board is connected to
sensor 1 terminal.
Troubleshoot
Reference Manual
May 2019
00809-0100-3412
86
Emerson.com/Rosemount
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...