9.7.3
Alarm relays don't work.
Verify the relays are properly wired.
9.8
Simulate inputs - chlorine
To check the performance of the transmitter, use a decade box and 1.5 V battery to
simulate the current from the sensor. The battery, which opposes the polarizing voltage, is
necessary to ensure that the sensor current has the correct sign.
Procedure
1. Disconnect the anode and cathode leads from terminals 1 and 2 on TB3 and
connect a decade box and 1.5 V battery as shown in
.
Figure 9-2: Simulating Chlorine
A. Anode shield
B. Anode
C. Cathode shield
D. Cathode
It is not necessary to disconnect the RTD leads.
2. Set the decade box to 2.8 MΩ.
3. Note the sensor current.
It should be about 500 nA. The actual value depends on the voltage of the battery.
To view the sensor current, go to the main display and press
DIAG
. Choose sensor 1.
The input current is the second line in the display.
4. Change the decade box resistance and verify that the correct current is shown.
Calculate current from the equation:
The voltage of a fresh 1.5 volt battery is about 1.6 volt (1600 mV).
Troubleshoot
Reference Manual
May 2019
00809-0100-3412
102
Emerson.com/Rosemount
Содержание Rosemount FCL
Страница 4: ...4 ...
Страница 10: ...Description and specifications Reference Manual May 2019 00809 0100 3412 10 Emerson com Rosemount ...
Страница 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Страница 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Страница 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Страница 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Страница 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Страница 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...