Mounting Dimensions for the Control and Power Supply Units
Figure 2-6:
A. Cover seal - Note that the mounting holes are outside the extent of the seal.
B. Cover
C. Base
D. 4 Holes for M6 mounting screws
E. Cable gland entry blanking page
F. Approx. 6 in. (150 mm) free space required below box for cables
G. Assembled box 4.3 in. (110 mm) deep
NOTICE
The unit mounting holes are located outside the seal. You do not need to seal the
mounting holes after installation or to remove the circuit boards from the unit prior to
mounting.
2.5.7
Power supply unit
Rosemount supplies adequate cableto locate the power supply unit up to 33 ft. (10 m)
from the source unit. You may use a maximum cable length of 82 ft. (25 m) to connect the
power supply unit to the receiver. Do not exceed the 82 ft. (25 m) maximum cable length.
Dimensions and mounting hole locations are identical to the control unit and are shown in
Install