Emerson Process Management GmbH & Co. OHG
1-16
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
05/2017
Tab. 1-4:
Oxygen - Standard and Enhanced Measurement Performance Specifications
1.6 Measurement Specifications
1
Related to full scale
2
Related to measuring value
3
From gas analyzer inlet at gas flow of 1.0 l/min
(electronic damping = 0 s)
4
Constant pressure and temperature
5
Range 0–10…200 ppm: ≤ 5 % (5 to 45 °C /
41 to 113 °F)
6
Pressure sensor is required
7
Limited to atmospheric if internal sample pump
8
Temperatures below 0 °C (-4 °F) with thermostat
control only
9
Thermost. controlled sensor: 35 °C (95 °F)
10
For ranges 0–5…100 %
and
flow 0.5…1.5 l/min
11
Optional thermostatically controlled sensor with
temperature 60 °C (140 °F)
12
Temperature variation:
≤
10 K per hour
13
No sudden pressure surge allowed
14
Special conditions apply to model XEFD
Oxygen Sensors
Paramagnetic (pO
2
)
Electrochemical (eO
2
)
Trace (tO
2
)
Standard Spec Enhanced Spec
Detection limit (4 σ)
1 4
≤
1 %
≤
0.5 %
≤
1 %
≤
1 %
Linearity
1 4
≤
1 %
≤
1 %
≤
1 %
Zero-point drift
1 4
≤
2 % per week
≤
1 % per week
≤
2 % per week
≤
1 % per week
Span (sensitivity) drift
1 4
≤
1 % per week
≤
0.5 % per week
≤
1 % per week
≤
1 % per week
Repeatability
1 4
≤
0.5 %
≤
1 %
≤
1 %
Response time (t
90
)
3
< 5 s
approx. 12 s
20 to 80 s
Permissible gas flow
0.2–1.5 l/min
0.2–1.5 l/min.
0.2–1.5 l/min.
Influence of gas flow
1
4
≤
2 %
10
≤
2 %
≤
2 %
Maximum gas pressure
7 14
≤ 1500 hPa abs. (≤
7 psig)
13
≤ 1500 hPa abs. (≤
7 psig)
≤ 1500 hPa abs. (≤
7 psig)
Influence of pressure
2
– At constant temperature
≤
0.10 % per hPa
≤
0.10 % per hPa
≤
0.10 % per hPa
– With pressure compensation
6
≤
0.01 % per hPa
≤
0.01 % per hPa
≤
0.01 % per hPa
Permissible ambient temperature
8
0(-20) to +50 °C (32 (4) to 122 °F) 5 to +45 °C (41 to 113 °F) 5 to +45 °C (41 to 113 °F)
Influence of temperature
1 12
(at constant pressure)
– On zero point
≤
1 % per 10 K
≤
0.5 % per 10 K
≤
1 % per 10 K
≤ 1 % per 10 K
5
– On span (sensitivity)
≤
1 % per 10 K
≤
1 % per 10 K
≤ 1 % per 10 K
5
Thermostat control
60 °C (140 °F)
11
none
none
9
Warm-up time
Approx. 50 minutes
-
Approx. 50 minutes
Note!
1 psi = 68.95 hPa
Note! Take care of the tO
2
sensor‘s documentation, providing important calibration instructions!
Hydrogen Sulfide (H
2
S)
Measurement range (sensor dependent)
0 to 50 ppm
0 to 200 ppm
0 to 2000 ppm
Overgas limit
200 ppm
500 ppm
10,000 ppm
Detection limit
1
< 0.2 %
Repeatability
1
< 2 %
Drift
1
< 1 % per month
Response time (t
90
)
< 35 s
Operating life
> 24 months
Sensor operating temperature
-30 to 50 °C (-22 °F to 122 °F)
Gas pressure range
800 to 1200 hPa
(-3.1 to 2.7 psig)
Gas humidity range (rel. humidity)
15 to 90 %
Thermostat control
none
Tab. 1-5:
H
2
S - Standard Measurement Performance Specifications
1
Related to full scale
Note
! These sensors require oxygen and moisture to work properly within given specifications!
Take care of the separate documentation accompanying the sensors!
Содержание Rosemount Analytical X-STREAM Enhanced Series
Страница 64: ...Emerson Process Management GmbH Co OHG 2 28 X STREAM XE Short Form Manual HASXEE SFM HS 05 2017 ...
Страница 94: ...Emerson Process Management GmbH Co OHG 3 30 X STREAM XE Short Form Manual HASXEE SFM HS 05 2017 ...
Страница 122: ...X STREAM XE Short Form Manual HASXEE SFM HS 05 2017 ...