Emerson Process Management GmbH & Co. OHG
S-10
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
Safety Instructions
GASES AND PREPARATION OF GASES
GASES HAZARDOUS TO HEALTH
Follow the safety precautions for all gases (sample and span gases) and
gas cylinders.
Before opening the gas lines, they must be purged with air or neutral gas (N2)
to avoid hazards from escaping toxic, flammable, explosive or hazardous
gases.
FLAMMABLE OR EXPLOSIVE GASES
The analyzers are not suitable to measure explosive gases. When supplying
flammable gases with concentrations of more than 25 % of the lower
explosion limit, we RECOMMEND implementing one or more additional
safety measures:
• purging the unit with inert gas
• stainless steel internal pipes
• flame arrestors on gas inlets and
outlets
• infallible measuring cells.