7.4.3 Procedure
1. Press PROG on the remote controller.
2. Press NEXT until the
tEMP
submenu appears. Press ENTER.
3. The screen displays the
tAUtO
(automatic temperature compensation) prompt.
Press
é
or
ê
to toggle between
On
and
OFF
. Press ENTER to save.
4. If you disable tAuto, the
tMAN
prompt appears. Use the arrow keys to change the
temperature to the desired value. To enter a negative number, press
è
or
ç
until no
digit is flashing. Then press
é
or
ê
to display the negative sign.
The temperature
entered in this step will be used in all measurements (oxygen, chlorine, ozone,
or pH), no matter what the process temperature is.
Press ENTER to save.
5. Press RESET to return to the process display.
PROGRAM
tAUtO
EXIT
ENTER
ON
PROGRAM
tMAn
EXIT
ENTER
0
25
.0
PROGRAM
tEMP
EXIT
NEXT ENTER
MODEL 5081-A
SECTION 7.0
PROGRAMMING