4.4
Recall factory trim
The recall factory trim commands allow the restoration of the as-shipped factory settings of
the sensor trim and analog output trim.
4.4.1
Recall factory trim—sensor trim
This command resets the transmitter sensor trim to the “as shipped” factory settings. The
recall factory trim—sensor trim command can be useful for recovering from an inadvertent
zero trim of an absolute pressure unit or inaccurate pressure source.
AMS Device Manger
1.
Right click on the device and select
Calibrate
, then
Recall Factory Trim
from the
menu
.
2.
Select
Next
after setting the control loop to manual.
3.
Select
Sensor trim
under
Trim to recall
and select
Next
.
4.
Select
Next
to acknowledge restoration of trim values is complete.
5.
Select
Next
to acknowledge the loop can be returned to automatic control.
6.
Select
Finish
to acknowledge the method is complete.
4.4.2
Recall factory trim—analog output
This command resets the transmitter analog output trim to the “as shipped” factory
settings. The recall factory trim—analog output command can be useful for recovering from
an inadvertent trim, incorrect Plant Standard or faulty meter.
AMS Device Manger
1.
Right click on the device and select
Calibrate
, then
Recall Factory Trim
from the
menu
.
2.
Select
Next
after setting the control loop to manual.
3.
Select
Analog output trim
under
Trim to recall
and select
Next
.
4.
Select
Next
to acknowledge restoration of trim values is complete.
5.
Select
Next
to acknowledge the loop can be returned to automatic control.
6.
Select
Finish
to acknowledge the method is complete.
Fast Keys
1, 2, 3, 4, 1
Fast Keys
1, 2, 3, 4, 2