The resulting potentials, measured with a clean platinum electrode and saturated
KCl/AgCL reference electrode, should be within ±20 millivolts of the value shown in
. Note solution temperature to ensure accurate interpretation of results.
The ORP value of saturated quinhydrone solution is not stable over long periods of
time. Therefore, make these standards fresh each time they are used.
Table 3-1: ORP of Saturated Quinhydrone Solution (Millivolts)
pH 4 solution
pH 7 solution
Temp ºC
20
25
30
20
25
30
mV
potential
268
264
260
94
87
80
5. Remove the sensor from the buffer, rinse, and install in the process.
Reference Manual
Calibration
00809-0100-3389
February 2019
Reference Manual
29
Содержание Rosemount 389
Страница 4: ...Contents Reference Manual February 2019 00809 0100 3389 4 Rosemount 389 389VP ...
Страница 10: ...Description and Specifications Reference Manual February 2019 00809 0100 3389 10 Rosemount 389 389VP ...
Страница 26: ...Install Reference Manual February 2019 00809 0100 3389 26 Rosemount 389 389VP ...
Страница 30: ...Calibration Reference Manual February 2019 00809 0100 3389 30 Rosemount 389 389VP ...
Страница 34: ...Maintenance Reference Manual February 2019 00809 0100 3389 34 Rosemount 389 389VP ...
Страница 38: ...Return of Material Reference Manual February 2019 00809 0100 3389 38 Rosemount 389 389VP ...
Страница 40: ...EC Declaration of Conformity Reference Manual February 2019 00809 0100 3389 40 Rosemount 389 389VP ...
Страница 41: ...Reference Manual EC Declaration of Conformity 00809 0100 3389 February 2019 Reference Manual 41 ...
Страница 42: ...EC Declaration of Conformity Reference Manual February 2019 00809 0100 3389 42 Rosemount 389 389VP ...
Страница 43: ...B China RoHS Table Reference Manual China RoHS Table 00809 0100 3389 February 2019 Reference Manual 43 ...
Страница 44: ...China RoHS Table Reference Manual February 2019 00809 0100 3389 44 Rosemount 389 389VP ...