Figure 2-4: Flow Through and Insertion Installations
A. Angle flow
B. Straight flow
C. Pipe "Y"
D. 1
½
-in. x 1-in. reducing bushing
E. 1
½
-in. pipe tee
F. 1
½
-in. pipe "Y"
G. flow
Note
Valves and fittings by others. Mount the sensor at least 10° from horizontal.
Note
Do not use large pipe wrenches to tighten the sensor into a flange or other type of
mounting.
2.3
Electrical installation
provide the guidelines for wiring Rosemount 389/389VP
sensors to various Rosemount transmitters.
1. If the cable needs to be extended, use a high quality four-conductor shielded
instrument cable available from Rosemount.
Note
If the cable is too long, loop up the excess cable. If the cable has to be shortened,
splice and terminate each conductor and make sure that the overall (outermost)
drain wire is not shorted out with either of the two inner drain wires (shields).
2. Signal cable should be run in a dedicated conduit (preferably an earth grounded
metallic conduit) and should be kept away from AC power lines. A spade lug kit is
provided in a plastic bag wrapped around the cable.
Reference Manual
Install
00809-0100-3389
February 2019
Reference Manual
15
Содержание Rosemount 389
Страница 4: ...Contents Reference Manual February 2019 00809 0100 3389 4 Rosemount 389 389VP ...
Страница 10: ...Description and Specifications Reference Manual February 2019 00809 0100 3389 10 Rosemount 389 389VP ...
Страница 26: ...Install Reference Manual February 2019 00809 0100 3389 26 Rosemount 389 389VP ...
Страница 30: ...Calibration Reference Manual February 2019 00809 0100 3389 30 Rosemount 389 389VP ...
Страница 34: ...Maintenance Reference Manual February 2019 00809 0100 3389 34 Rosemount 389 389VP ...
Страница 38: ...Return of Material Reference Manual February 2019 00809 0100 3389 38 Rosemount 389 389VP ...
Страница 40: ...EC Declaration of Conformity Reference Manual February 2019 00809 0100 3389 40 Rosemount 389 389VP ...
Страница 41: ...Reference Manual EC Declaration of Conformity 00809 0100 3389 February 2019 Reference Manual 41 ...
Страница 42: ...EC Declaration of Conformity Reference Manual February 2019 00809 0100 3389 42 Rosemount 389 389VP ...
Страница 43: ...B China RoHS Table Reference Manual China RoHS Table 00809 0100 3389 February 2019 Reference Manual 43 ...
Страница 44: ...China RoHS Table Reference Manual February 2019 00809 0100 3389 44 Rosemount 389 389VP ...