Figure 2-3: Submersion Installations
4. Run the sensor cable through a protective conduit for isolation from electrical
interference or physical abuse from the process.
5. Install the sensor within 80° of vertical, with the electrode facing down.
Do not run the sensor's cable with power or control wiring.
2.2.2
Flow through and insertion mounting
Rosemount 389 and 389VP sensors also have a 1-in. MNPT process at the front of the
sensor for mounting into a 1
¼
configurations.
Install
Reference Manual
February 2019
00809-0100-3389
14
Rosemount 389/389VP
Содержание Rosemount 389
Страница 4: ...Contents Reference Manual February 2019 00809 0100 3389 4 Rosemount 389 389VP ...
Страница 10: ...Description and Specifications Reference Manual February 2019 00809 0100 3389 10 Rosemount 389 389VP ...
Страница 26: ...Install Reference Manual February 2019 00809 0100 3389 26 Rosemount 389 389VP ...
Страница 30: ...Calibration Reference Manual February 2019 00809 0100 3389 30 Rosemount 389 389VP ...
Страница 34: ...Maintenance Reference Manual February 2019 00809 0100 3389 34 Rosemount 389 389VP ...
Страница 38: ...Return of Material Reference Manual February 2019 00809 0100 3389 38 Rosemount 389 389VP ...
Страница 40: ...EC Declaration of Conformity Reference Manual February 2019 00809 0100 3389 40 Rosemount 389 389VP ...
Страница 41: ...Reference Manual EC Declaration of Conformity 00809 0100 3389 February 2019 Reference Manual 41 ...
Страница 42: ...EC Declaration of Conformity Reference Manual February 2019 00809 0100 3389 42 Rosemount 389 389VP ...
Страница 43: ...B China RoHS Table Reference Manual China RoHS Table 00809 0100 3389 February 2019 Reference Manual 43 ...
Страница 44: ...China RoHS Table Reference Manual February 2019 00809 0100 3389 44 Rosemount 389 389VP ...