Quick Installation Guide
00825-0100-4774, Rev DA
April 2010
Rosemount 3051
8
S
TEP
3: C
ONSIDER
H
OUSING
R
OTATION
To improve field access to wiring or to better view the optional LCD display:
S
TEP
4: C
ONNECT
THE
W
IRING
AND
P
OWER
U
P
1. Loosen the housing rotation set screw.
2. First rotate the housing clockwise to the desired location. If
the desired location cannot be achieved due to thread limit,
rotate the housing counter clockwise to the desired location
(up to 360° from thread limit).
3. Retighten the housing rotation set screw.
Use of ordinary copper wire of sufficient size to
ensure that the voltage across the transmitter
power terminals does not drop below 9 vdc. To
power the transmitter, connect the power leads to
the terminals indicated on the terminal block label.
The power terminals are polarity insensitive, which
means the electrical polarity of the power leads
does not matter when connecting to the power
terminals. When wiring to the screw terminals, the
use of crimped legs is recommended. Tighten the
terminal screws to ensure adequate contact. No
additional power is needed.
Housing Rotation Set Screw
(5/64-inch)
Ground
Terminal
Power
Terminals
“NC” is a No Connect terminal (do not use)
4774RevDA.fm Page 8 Thursday, April 22, 2010 10:26 AM
Содержание Rosemount 3051
Страница 17: ...Quick Installation Guide 00825 0100 4774 Rev DA April 2010 Rosemount 3051 17 doc ...
Страница 18: ...Quick Installation Guide 00825 0100 4774 Rev DA April 2010 Rosemount 3051 18 ...
Страница 19: ...Quick Installation Guide 00825 0100 4774 Rev DA April 2010 Rosemount 3051 19 ...
Страница 20: ...Quick Installation Guide 00825 0100 4774 Rev DA April 2010 Rosemount 3051 20 ...