5.3
Check the probe wiring
Prerequisites
The power supply to the level switch must be switched off.
Procedure
1. Clean away any deposits on the probe.
2.
Take out the electronic board and disconnect internal wires.
3.
Check the orange, yellow, and green/yellow wires with a
multimeter.
A
B
C
D
E
A. Orange (probe) and yellow (shield)
B. Green/yellow (ground)
C. Ground
D. Shield
E. Probe
Less than 5 Ohm must be present between:
•
Orange wire and probe
•
Yellow wire and shield
•
Green/yellow wire and ground
More than 1 MΩ resistance must be present between:
•
Orange and yellow wires
•
Orange and green/yellow wires
If other values are present, the wiring of the probe is incorrect or
defective.
October 2020
Quick Start Guide
Quick Start Guide
43