Figure 2-4: Correct and Incorrect Mounting (Rope Version)
> 1.8 (50)
> 1.8 (50)
A
A
OK
OK
B
C
D
E
A. Active probe
B. Correct installation: The socket height is short
C. Correct installation: The inactive length is correctly used within a long
socket
D. Incorrect installation: The active probe is inside the socket
E. Incorrect installation: The active probe is covered by a build-up of
material and may not be detecting the true level
October 2020