Transmitter Programming

39

Instruction Manual

Section 7: Transmitter programming

LIQ-MAN-1066

April 2017

7.4.4

7.5

7.5.1

7.5.2

Procedure: Ranging the Current Outputs 

The  adjacent  screen  will  appear  under  Program/Output/Range.

Enter  a  value  for  4  mA  and  20  mA  for  each  output.  Follow  the

menu screens in 

Figure 7-1

to assign values to the outputs.

Setting a Security Code

Purpose

The security codes prevent accidental or unwanted changes to program settings, displays, and

calibration. The 1066 has two levels of security code to control access and use of the instrument

to different types of users. The two levels of security are: 

all:

This  is  the  Supervisory  security  level.  It  allows  access  to  all  menu  functions,  including

Programming, Calibration, Hold and Display.

Calibration/hold:

This is the operator or technician level menu. It allows access to only cali-

bration and Hold of the current outputs. 

Procedure

1. Press MENU. The main menu screen appears. Choose 

program

.

2. Scroll down to Security.  Select Security.  
3. The security entry screen appears.  Enter a three digit security code for each of the desired

security levels. The security code takes effect two minutes after the last key stroke. Record the

security code(s) for future access and communication to operators or technicians as needed. 

4. The display returns to the security menu screen. Press EXIT to return to the previous screen.

To return to the main display, press MENU followed by EXIT.

Figure 7-2

displays the security code screens.   

1.234µS/cm    25.0 ºC

Output Range

O1 SN 4mA:    0.000 µS/cm

O1 SN 20mA:  20.00 µS/cm

O2 SN 4mA:    0 °C

O2 SN 20mA:  100 °C

     1.234µS/cm    25.0ºC

Program

Outputs
Measurement
Temperature

Security
Diagnostic Setup
Reset Analyzer

Outputs

Program

Range

Configure

Simulate

     1.234µS/cm    25.0ºC

Output Configure

Output1 
Output2 

      1.234µS/cm    25.0ºC

Outputs

Range 
Configure
Simulate

     1.234µS/cm    25.0ºC

Output Range

O1 SN 4mA:              0.000
O1 SN 20mA:            20.00
O2 SN 4mA:              0ºC
S2 20mA:                  100ºC

     1.234µS/cm    25.0ºC

Simulate

Output 1
Output 2

     1.234µS/cm    25.0ºC

Output Range

OM 4mA:   1.000 
OM 20mA:  1.000% 
OM 4mA:    1.000%
OM 20mA:  7.0 pH

     1.234µS/cm    25.0ºC

OutputN Hold At

12.00mA

     1.234µS/cm    25.0ºC

OutputM Configure

Assign:            S1 Meas
Range:             4-20mA
Scale:                 Linear
Dampening:          0sec

MAIN MENU

     1.234µS/cm    25.0ºC

OutputM Assign

S1 Measurement
S1 Temperature

     1.234µS/cm    25.0ºC

OutputM Range

4-20mA
0-20mA

     1.234µS/cm    25.0ºC

Scale

Linear
Log

FIGuRE 7-1. Configuring and

Ranging the Current Outputs

Содержание Rosemount 1066

Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...

Страница 2: ......

Страница 3: ... performance use qualified personnel to install operate update program and maintain the product When replacement parts are required ensure that qualified people use replacement parts specified by Rosemount Unauthorized parts and procedures can affect the product s performance and place the safe operation of your process at risk Look alike substitutions may result in fire electrical hazards or impr...

Страница 4: ...rds and Intrinsically Safe installation drawings and update CE certificates Added FM temperature specifications to Non Incendive Hazardous Location Approval F 9 2013 Added Section 10 HART Communications G 11 2014 Changed agency water exposure testing description to Type H 05 2015 FM approvals updated J 04 2017 Updated the Address and Emerson logo Also updated the FM CSA installation drawings and C...

Страница 5: ...commended Sensors for Conductivity 9 2 6 Chlorine Codes L 9 2 6 1 Free and Total Chlorine 9 2 6 2 Performance Specifications 9 2 6 3 Recommended Sensors 9 2 6 4 Monochloromine 9 2 6 5 Performance Specifications 10 2 6 6 Recommended Sensors 10 2 7 Dissolved Oxygen Codes DO 10 2 7 1 Performance Specification 10 2 7 2 Recommended Sensors 10 2 8 Dissolved Oxygen Codes DO 10 2 8 1 Performance Specifica...

Страница 6: ... General 37 7 2 Changing the Startup Settings 37 7 2 1 Purpose 37 7 2 2 Procedure 38 7 3 Choosing Temperature Units and Automatic Manual Temperature Compensation 38 7 3 1 Purpose 38 7 4 Configuring and Ranging Current Outputs 38 7 4 1 Purpose 38 7 4 2 Definitions 38 7 4 3 Procedure Configure Outputs 38 7 4 4 Procedure Ranging the Current Outputs 38 7 5 Setting a Security Code 38 7 5 1 Purpose 39 7...

Страница 7: ...4 1 Description 47 8 4 2 Sensor Type 47 8 4 3 Measure 48 8 4 4 Range 48 8 4 5 Cell Constant 48 8 4 6 RTD Offset 48 8 4 7 RTD Slope 48 8 4 8 Temp Comp 48 8 4 9 Slope 49 8 4 10 Reference Temp 49 8 4 11 Filter 49 8 4 12 Custom Setup 49 8 4 13 Cal Factor 49 8 5 Toroidal Conductivity Measurement Programming 50 8 5 1 Description 50 8 5 2 Sensor Type 50 8 5 3 Measure 51 8 5 4 Range 51 8 5 5 Cell Constant...

Страница 8: ... Units 58 8 7 3 Partial Press 58 8 7 4 Salinity 58 8 7 5 Filter 58 8 7 6 Pressure Units 58 8 8 Ozone Measurement Programming 59 8 8 1 Units 59 8 8 2 Filter 59 8 8 3 Resolution 59 Section 9 Calibration 9 1 Introduction 67 9 2 Calibration 67 9 2 1 Auto Calibration 68 9 2 2 Manual Calibration pH 68 9 2 3 Entering a Known Slope Value pH 68 9 2 4 Standardization pH 69 9 2 5 SMART sensor auto calibratio...

Страница 9: ...9 7 3 Calibrating the Sensor Against A Standard Instrument in process cal 83 9 8 Calibration Ozone 83 9 8 1 Zeroing the Sensor 84 9 8 2 In Process Calibration 84 9 9 Calibrating Temperature 85 9 9 1 Calibration 85 Section 10 haRT Communications 10 1 Introduction 93 10 2 Physical Installation and Configuration 94 10 3 Measurements Available via HART 96 10 4 Diagnostics Available via HART 96 10 5 HA...

Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 11: ... desired language and press ENTER 6 Choose measurement and press ENTER a For pH choose preamplifier location Select Analyzer to use the integral preamplifier in the transmitter select Sensor J Box if your sensor is SMART or has an integral preamplifier or if you are using a remote preamplifier located in a junction box 5 If applicable choose units of measurement 6 For contacting and toroidal condu...

Страница 12: ...2 Description and Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 13: ...red The instrument prompts the user to configure the sensor loop in a few quick steps for immediate commis sioning user help Screens Fault and warning messages include help screens similar to PlantWeb alerts that provide useful troubleshooting tips to the user These on screen instructions are intuitive and easy to use Diagnostics The transmitter continuously monitors itself and the sensor for prob...

Страница 14: ... IEC 60079 0 2011 Edition 6 0 I EC 60079 11 2011 06 Edition 6 0 FM 3600 2011 3610 2010 3611 2004 3810 2005 IEC 60529 2004 ANSI ISA 60079 0 2009 ANSI ISA 60079 11 2009 2 2 hazardous location approvals Intrinsic Safety with appropriate safety barrier Class I II III Div 1 Groups A G T4 Tamb 20 C to 65 C Enclosure 4X IP66 For Intrincically Safe Installation see drawing 1400669 1180 II 1 G Baseefa11ATE...

Страница 15: ...us the external load resistor required for HART communications 250 Ω minimum Minimum power supply voltage is 12 7 Vdc Maximum power supply voltage is 42 4 Vdc 30 Vdc for intrinsically safe operation The graph shows the supply voltage required to maintain 12 Vdc upper line and 30 Vdc lower line at the transmitter terminals when the current is 22 mA analog Outputs Two wire loop powered Output 1 only...

Страница 16: ...n is menu selectable Change in pH due to process temperature can be compensated using a programmable temperature coefficient Performance Specifications Transmitter pH input Measurement Range ph 0 to 14 pH accuracy 0 01 pH Buffer recognition NIST DIN 19266 JIS 8802 and BSI Input filter Time constant 1 999 sec default 4 sec Response time 5 seconds to 95 of final reading Recommended Sensors for ph Al...

Страница 17: ...cting conductivity sensors refer to the product data sheets Note The 410VP 4 electrode high range conductivity sensor is compatible with the 1066 Performance Specifications Temperature specifications Input filter Time constant 1 999 sec default 2 sec Response time 3 seconds to 95 of final reading using the default input filter Salinity Uses Practical Salinity Scale Total Dissolved Solids Calculate...

Страница 18: ...mum results Two temperature compensation options are available manual slope X C and neutral salt dilute sodium chloride Temperature compensa tion can be disabled allowing the transmitter to display raw conductivity For more information con cerning use of the toroidal conductivity sensors refer to the product data sheets Performance Specifications Temperature specifications Repeatability 0 25 5 µS ...

Страница 19: ...erning the use and operation of the amperometric chlorine sensors and the TCL measurement system refer to the product data sheets Performance Specifications Resolution 0 001 ppm or 0 01 ppm selectable Input Range 0nA 100 µA automatic ph correction for Free Chlorine user selectable for code CL 6 0 to 10 0 pH Temperature compensation Automatic via RTD or manual 0 50 C Input filter Time constant 1 99...

Страница 20: ...ensors refer to the product data sheets Performance Specifications Resolution 0 01 ppm 0 1 ppb for 499A TrDO sensor when O2 1 00 ppm 0 1 Input Range 0nA 100μA Temperature Compensation Automatic via RTD or manual 0 50 C Input filter Time constant 1 999 sec default 5 sec Response time 6 seconds to 95 of final reading Recommended Sensors Rosemount amperometric membrane and steam sterilizable sensors ...

Страница 21: ...s an option Communication with the 1066 is through Local keypad interface 475 HART and FOUNDATION fieldbus Communicator HART protocol version 7 FOUNDATION fieldbus AMS Asset Management Solutions Aware SMART Wireless THUM Adapter 2 9 TaBlE 2 3 Ordering Information Description 1066 ph ORp Conductivity Chlorine Oxygen and Ozone 2 Wire Transmitter Measurement P pH ORP C Contacting Conductivity T Toroi...

Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 23: ...ansmitter case has six 1 2 inch PG13 5 conduit openings Use separate conduit openings for the power output cable the sensor cable and the other the sensor cable as needed pH input for free chlorine with continuous pH correction 6 Use weathertight cable glands to keep moisture out to the transmitter If conduit is used plug and seal the connections at the transmitter housing to prevent moisture from...

Страница 24: ...14 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 FIGuRE 3 1 panel Mounting Dimensions ...

Страница 25: ...Installation 15 Instruction Manual Section 3 Installation LIQ MAN 1066 April 2017 FIGuRE 3 2 pipe and wall mounting dimensions Mounting bracket pN 23820 00 ...

Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 27: ...upply must be able to cover the voltage drop on the cable as well as the load resistor 250 Ω minimum required for HART communications The maximum power supply voltage is 42 0 Vdc For intrinsically safe installations the maximum power supply voltage is 30 0 Vdc The graph shows load and power supply requirements The upper line is the power supply voltage needed to provide 12 7 Vdc at the transmitter...

Страница 28: ...wer supply 2 Use a metal cable gland and be sure the shield makes good electrical contact with the gland 3 Use the metal backing plate when attaching the gland to transmitter enclosure The power signal cable can also be enclosed in an earth grounded metal conduit Do not run power supply signal wiring in the same conduit or cable tray with loop power lines Keep power supply signal wiring at least 6...

Страница 29: ...position on the main board using the lead markings positive negative on the board OUTPUT 2 TB5 TB3 TB2 TB4 TB1 24V GND GND 24V THUM ANODE CATHODE RTN SNS RTD IN V V REF SHLD SENSOR WIRING TB7 TB6 _ GND SOL SHLD pH OUTPUT1 LOOP PWR 4 20mA 24VDC RETURN 4 20mA 24VDC INSTALL PLUGS IN ALL OTHER OPENINGS AS NEEDED INNER ENCLOSURE HINGE SIDE OF FRONT PANEL 1 TB7 OUTPUT 2 REQUIRES EXTERNALDC POWER 2 TB6 T...

Страница 30: ...h grounded metal conduit Do not run power supply signal wiring in the same conduit or cable tray with loop power lines Keep power supply signal wiring at least 6 ft 2 m away from heavy electrical equipment FOuNDaTION Fieldbus Figure 4 4 shows a 1066PFF being used to measure and control pH and chlorine levels in drinking water The figure also shows three ways in which Fieldbus communication can be ...

Страница 31: ...SMART pH sensors can be wired to the 1066 using integral cable SMART sensors or compatible VP8 pH cables After wiring the sensor leads carefully take up the excess sensor cable through the cable gland Keep sensor and output signal wiring separate from loop power wiring Do not run sensor and power wiring in the same conduit or close together in a cable tray 4 4 FIGuRE 4 6 ph ORp sensor wiring to th...

Страница 32: ...22 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 DWG NO 40106615 FIGuRE 4 7 Contacting and Toroidal Conductivity sensor wiring to the 1066 circuit board ...

Страница 33: ... FOLLOW RECOMMENDED ORDER ANODE CATHODE 1 TB5 ANODE CATHODE RETURN SENSE 2 TB3 RTD SOLUTION GROUND 3 TB2 SOLUTION GROUND RTD IN NO CONNECTION NO CONNECTION NOTE A TB1 TB4 TB6 AND TB7 NOT USED FOR OXYGEN AND OZONE SENSOR WIRING B TB1 TB2 AND TB4 MAY BE USED FOR pH SENSOR WIRING IF FREE CHLORINE MEASUREMENT REQUIRES LIVE pH INPUT 1066 CIRCUIT BOARD ASSY 24406 xx DWG NO 40106611 FIGuRE 4 8 Chlorine o...

Страница 34: ...UM OUTPUT1 LOOP PWR 4 20mA 24VDC RETURN 4 20mA 24VDC 1 TB7 OUTPUT 2 REQUIRES EXTERNAL DC POWER 2 TB6 THUM TERMINAL IS USED ONLY FOR WIRELESS THUM ADAPTOR INSTALLATIONS 250 OHM RESISTOR IS PRE INSTALLED IN CIRCUIT 3 SPLICE CONNECTOR PROVIDED BY END USER INSTALL PLUGS IN ALL OTHER OPENINGS AS NEEDED INNER ENCLOSURE HINGE SIDE OF FRONT PANEL 1 2 WIRELESS THUM ADAPTOR 1066 HART CIRCUIT BOARD pH CL DO ...

Страница 35: ... GND SOL SHLD pH OUTPUT1 LOOP PWR 4 20mA 24VDC RETURN 4 20mA 24VDC INSTALL PLUGS IN ALL OTHER OPENINGS AS NEEDED INNER ENCLOSURE HINGE SIDE OF FRONT PANEL 1 TB7 OUTPUT 2 REQUIRES EXTERNALDC POWER 2 TB6 THUM TERMINAL IS USED ONLY FOR WIRELESS THUM ADAPTOR INSTALLATIONS 1 2 1066 HART CIRCUIT BOARD pH CL DO OZ ASSY 24406 xx HINGED PANEL DWG NO 40106613 FIGuRE 4 10 haRT loop power Wiring ...

Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 37: ... TERMINALS TB7 1 2 1066 HT LOOP POWER SIGNAL TERMINALS TB6 1 2 3 Vmax VDC 30 30 30 17 5 Imax mA 200 200 300 380 Pmax W 0 9 0 9 1 3 5 32 Ci nF 0 0 0 0 Li mH 0 0 0 0 TABLE III OUTPUT PARAMETERS Io Uo Po A B TABLE II FOR 1066 P CL DO OZ D C Ca µF La mH OUTPUT PARAMETERS GAS GROUPS MODEL 1066 TB1 1 THRU 12 1 44 11 88 V 153 4 mA 231 mW 1 51 9 39 6 04 38 5 12 08 TABLES I AND II ARE FOR pH CHLORINE DISSO...

Страница 38: ...DAPTER ALTERNATE POWER CONNECTION IF SMART THUM WIRELESS ADAPTER IS USED SENSOR AND SECOND ANALOG OUTPUT CONNECTION UNCHANGED FROM ABOVE ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA SEE NOTE 3 AND TABLE III AMPEROMETRIC SENSOR CSA APPROVED DEVICE OR SIMPLE APPARATUS 1066 CL DO OZ ONLY EITHER OR BOTH MAY BE INSTALLED OPTIONAL CSA APPROVED PREAMP...

Страница 39: ...THAN 200 FEET OF ATTACHED CABLE UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY 24 VDC TYPICAL 17 5 VDC MAX FOR FISCO OPTION SAFETY BARRIER SEE NOTES 2 8 FOR FISCO SEE NOTES 2 8 9 10 FOR ALL OTHER OPTIONS LOAD UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY 24 VDC TYPICAL 17 5 VDC MAX FOR FISCO OPTION SAFETY BARRIER SEE NOTES 2 8 FOR FISCO SEE NOTES 2 8 9 10 FOR ALL OTHER OPTIONS...

Страница 40: ...NECTED TO THE APPARATUS REF TABLES I II AND III 2 THE MODEL 1066 C T HAS SYSTEM APPROVAL FOR USE WITH MODELS 222 225 226 228 TOROIDAL SENSORS OR 140 141 142 150 400 401 402 1066 MODELS WITH P L D Z OPTIONS HAVE OUTPUT ENTITY PARAMETERS WHICH ALLOW THE USE OF VARIOUS SENSORS WHICH MAY TO THE SENSOR TERMINALS DO NOT EXCEED THE VALUES SPECIFIED IN TABLE I WHERE 1 ANY SINGLE SHUNT ZENER DIODE SAFETY B...

Страница 41: ...OTES 10 11 UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR IS 24V TYPICAL LOAD 1066 CL DO OZ ONLY HAZARDOUS AREA IS CLASS I GRPS A D CLASS II GRPS E G CLASS III NI CLASS I DIV 2 GRPS A D CLASS II DIV 2 GRPS E G WARNING SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2 WARNING TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES DISCONNECT POWER BEFORE SERVICING NO...

Страница 42: ...MOSPHERES DISCONNECT POWER BEFORE SERVICING IS CLASS I GRPS A D CLASS II GRPS E G CLASS III NI CLASS I DIV 2 GRPS A D CLASS II DIV 2 GRPS E G OPTIONAL FOR FIELDBUS FISCO INSTALLATIONS ANY FM APPROVED TERMINATOR ANY FM APPROVED FF FISCO DEVICE ANY FM APPROVED FF FISCO DEVICE MULTIPLE DEVICES NUMBER IS LIMITED BY THE REQUIREMENT TO MEET ALL OTHER FIELDBUS I S FISCO REQUIREMENTS FOR THE NETWORK ROSEM...

Страница 43: ...ypad Function Keys The MENu key is used to access menus for programming and calibrating the instrument Four top level menu items appear when pressing the MENU key Calibrate calibrate the attached sensor and analog output s Hold Suspend current output s Program Program outputs measurement temperature security and reset Display Program display format language warnings and contrast Pressing MENu from...

Страница 44: ...splay quadrants at the bottom half of the screen All four secondary value positions can be programmed by the user to any displayable parameter available Fault and Warning Banner If the transmitter detects a problem with itself or the sensor the word Fault or Warning will appear at the bottom of the display A fault requires immediate attention A warning indicates a problem atic condition or an impe...

Страница 45: ...rn to a previous menu level or to enable the main live display press the EXIT key repeatedly To return immediately to the main display from any menu level simply press MENU then EXIT The selection keys have the following functions The Up key above ENTER increments numerical values moves the decimal place one place to the right or selects units of measurement The Down key below ENTER decrements num...

Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 47: ...en In Gas Ozone OZ Ozone ppm mg L ppb μg L General Typical programming steps include the following listed procedures Each of these programming functions are easily and quickly accomplished using the intuitive menu system Changing the measurement type measurement units and temperature units Choose temperature units and manual or automatic temperature compensation mode Configure and assign values to...

Страница 48: ...utputs This section provides a guide for configuring and ranging the outputs ALWAYS CONFIGURE THE OUTPUTS FIRST Definitions 1 CURRENT OUTPUTS The transmitter provides a continuous output current 4 20 mA directly proportional to the process variable or temperature The low and high current out puts can be set to any value 2 ASSIGNING OUTPUTS Assign a measurement or temperature to Output 1 or Output ...

Страница 49: ...rity levels The security code takes effect two minutes after the last key stroke Record the security code s for future access and communication to operators or technicians as needed 4 The display returns to the security menu screen Press EXIT to return to the previous screen To return to the main display press MENU followed by EXIT Figure 7 2 displays the security code screens 1 234µS cm 25 0 ºC O...

Страница 50: ...ecurity code for the appropriate security level 3 If the entry is correct the appropriate menu screen appears If the entry is incorrect the Invalid Code screen appears The Enter Security Code screen reappears after 2 seconds using hold Purpose The transmitter output is always proportional to measured value To prevent improper operation of systems or pumps that are controlled directly by the curren...

Страница 51: ...ure 7 3 Resetting Factory Default Settings Purpose This section describes how to restore factory calibration and default values The process also clears all fault messages and returns the display to the first Quick Start screen The 1066 offers three options for resetting factory defaults 1 Reset all settings to factory defaults 2 Reset sensor calibration data only 3 Reset analog output settings onl...

Страница 52: ...Section 7 Transmitter programming Instruction Manual April 2017 LIQ MAN 1066 42 Transmitter Programming FIGuRE 7 4 Resetting Factory Default Settings ...

Страница 53: ...fining measurement display resolution pH and amperometric 6 Defining measurement display units all sections 7 Adjusting the input filter to control display and output reading variability or noise all sections 8 Selecting a measurement range conductivity see Section 8 4 8 5 9 Entering a cell constant for a contacting or toroidal sensor see Section 8 4 8 5 10 Creating an application specific concent...

Страница 54: ...gram any function scroll to the desired item and press ENTER The following sub sections provide you with the initial display screen that appears for each configuration function Use the flow diagram for pH programming and the 1066 live screen prompts for each function to complete configuration and programming Measurement The display screen for selecting the measurement is shown The default value is...

Страница 55: ...default value is displayed in bold type Refer to the pH ORP Programming flow diagram to complete this function Reference Impedance The display screen for selecting Low or High Reference impedance is shown The default value is displayed in bold type Refer to the pH ORP Programming flow diagram to complete this function ORp Measurement programming The section describes how to configure the 1066 tran...

Страница 56: ...the pH ORP Programming flow diagram to complete this function Preamp The display screen for identifying the Preamp location is shown The default value is displayed in bold type Refer to the pH ORP Programming flow diagram to complete this function Filter The display screen for entering the input filter value in seconds is shown The default value is displayed in bold type Refer to the pH ORP Progra...

Страница 57: ... the Cal Factor for 4 Electrode sensors from the sensor tag To configure the contacting conductivity 1 Press MENU 2 Scroll down to Program Press ENTER 3 Scroll down to Measurement Press ENTER The adjacent screen format will appear factory defaults are shown To program any displayed function scroll to the desired item and press ENTER The following sub sections provide you with the initial display s...

Страница 58: ...or the sensor is shown The default value is displayed in bold type Refer to the contacting conductivity Programming flow diagram to complete this function RTD Slope The display screen for entering the RTD slope for the sensor is shown The default value is displayed in bold type Refer to the contacting conductivity Programming flow diagram to complete this function Temp Comp The display screen for ...

Страница 59: ... the determination of a custom curve fit to the entered data by displaying this screen If the custom curve fit is not completed or is unsuccessful the display will read as follows and the screen will return to the beginning custom curve screen Cal Factor Upon initial installation and power up if 4 electrode was selected for the sensor type in the Quick Start menus the user enters a Cell Constant a...

Страница 60: ...figure toroidal conductivity 1 Press MENU 2 Scroll down to Program Press ENTER 3 Scroll down to Measurement Press ENTER The adjacent screen format will appear factory defaults are shown To program any displayed function scroll to the desired item and press ENTER The following sub sections provide you with the initial display screen that appears for each configuration function Use the flow diagram ...

Страница 61: ...tion 9 5 2 zeroing the Instrument and Section 9 5 3 Calibrating the Sensor in a Conductivity Standard Cell Constant The display screen for entering a cell Constant for the sensor is shown The default value is displayed in bold type Refer to the toroidal conductivity Programming flow diagram to complete this function NOTE When manually changing ranges the Cell Constant may change through the calibr...

Страница 62: ...relative to the actual process value In these cases it is recommended to increase the input filter setting above the default value of 2 sec to suppress the effect of noise Custom Setup The display screens for creating custom curves for converting conductivity to concentration is shown Refer to the toroidal conductivity Programming flow diagram to complete this function When the custom curve data e...

Страница 63: ...its ppm Select units ppm or mg L 3 Resolution 0 001 Select display resolution 0 01 or 0 001 4 Free Cl Correct Live Select Live Continuous pH correction or Manual 5 Manual pH 7 00 pH For Manual pH correction enter the pH value 6 Filter 5sec Override the default input filter enter 0 999 seconds To configure chlorine for free chlorine 1 Press MENU 2 Scroll down to Program Press ENTER 3 Scroll down to...

Страница 64: ...rrection or Manual pH correction is shown The default value is displayed in bold type Refer to the Chlorine Programming flow diagram to complete this function Manual pH Correction The display screen for manually entering the pH value of the measured process liquid is shown The default value is displayed in bold type Refer to the Chlorine Programming flow diagram to complete this function Resolutio...

Страница 65: ... sections provide you with the initial display screen that appears for each configuration function Use the flow diagram for chlorine programming at the end of Sec 6 and the 1066 live screen prompts for each function to complete configuration and programming Measure The display screen for selecting the measurement is shown The default value is displayed in bold type Refer to the chlorine Programmin...

Страница 66: ...o Measurement Press ENTER The following screen format will appear factory defaults are shown To program any displayed function scroll to the desired item and press ENTER The following sub sections provide you with the initial display screen that appears for each configuration function Use the flow diagram for chlorine programming at the end of Sec 8 and the 1066 live screen prompts for each functi...

Страница 67: ...s 3 Units ppm Select ppm mg L ppb g L Sat O2 Gas ppm Oxygen Gas 4 Pressure Units bar Select pressure units mm Hg in Hg Atm kPa mbar bar 5 Salinity 00 0 Enter Salinity as 6 Filter 5 sec Override the default input filter enter 0 999 seconds 7 Partial Press mmHg Select mm Hg in Hg atm kPa mbar or bar for Partial pressure To configure Oxygen 1 Press MENU 2 Scroll down to Program Press ENTER 3 Scroll d...

Страница 68: ...red is shown The default value is dis played in bold type Refer to the Oxygen Programming flow diagram to complete this function Enter Salinity as Filter The display screen for entering the input filter value in seconds is shown The default value is displayed in bold type Refer to the Oxygen Programming flow diagram to complete this function Pressure Units The display screen for selecting pressure...

Страница 69: ...ram for ozone programming at the end of Sec 6 and the 1066 live screen prompts for each function to complete configuration and programming Note Ozone measurement boards are detected automatically by the analyzer No measurement selection is necessary Units The display screen for selecting measurement units is shown The default value is displayed in bold type Refer to the Ozone Programming flow dia ...

Страница 70: ...FIGuRE 8 1 Configuring ph ORp Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 60 Programming Measurements ...

Страница 71: ...DS Salinity 1 234µS cm 25 0ºC SN RTD Offset 0 00 C 1 234µS cm 25 0ºC SN RTD Slope 2 00 ºC 1 234µS cm 25 0ºC SN Temp Comp Slope Neutral Salt Cation Raw 1 234µS cm 25 0ºC SN Slope 2 00 ºC 1 234µS cm 25 0ºC SN Ref Temp 25 0ºC normal 25 0ºC 1 234µS cm 25 0ºC SN Custom Curve Configure Enter Data Points Calculate Curve 1 234µS cm 25 0ºC SN Custom Config Units ppm of Points 2 Ref Temp 25 0 ºC Slope 2 00 ...

Страница 72: ...234µS cm 25 0ºC SN RTD Slope 2 00 ºC 1 234µS cm 25 0ºC SN Temp Comp Slope Neutral Salt Raw 1 234µS cm 25 0ºC SN Slope 2 000 ºC 1 234µS cm 25 0ºC SN Ref Temp 25 0ºC normal 25 0ºC 1 234µS cm 25 0ºC SN Custom Curve Configure Enter Data Points Calculate Curve 1 234µS cm 25 0ºC SN Custom Config Units ppm of Points 2 Ref Temp 25 0 ºC slope 2 00 C 1 234µS cm 25 0ºC SN Units ppm mg L g L 1 234µS cm 25 0ºC...

Страница 73: ...r Cal Measurement Oxygen 1 234µS cm 25 0ºC SN Units ppm mg L ppb µg L Saturation Partial Pressure Oxygen In Gas ppm Oxygen In Gas 1 234µS cm 25 0ºC SN Salinity 00 0 1 234µS cm 25 0ºC SN Type Water Waste Trace Oxygen BioRx Rosemount BioRx Other Brewing Oxygen In Gas 1 234µS cm 25 0ºC SN Input filter 05 sec 1 234µS cm 25 0ºC SN Partial Press mm Hg in Hg atm kPa mbar bar 1 234µS cm 25 0ºC Presssure U...

Страница 74: ... pH Resolution 0 001 Measurement Chlorine 1 234µS cm 25 0ºC SN Units ppm mg L 1 234µS cm 25 0ºC SN Free Cl pH Correction Live Continuous Manual 1 234µS cm 25 0ºC SN Measurement Free Chlorine Total Chlorine Monochloramine 1 234µS cm 25 0ºC SN Manual pH 07 00 pH 1 234µS cm 25 0ºC SN Input filter 05 sec 1 234µS cm 25 0ºC SN Resolution 0 001 0 01 Section 8 programming Measurements Instruction Manual A...

Страница 75: ...S cm 25 0ºC SN Configure Units ppm Filter 5sec Resolution 0 001 Measurement Ozone 1 234µS cm 25 0ºC SN Input filter 05 sec 1 234µS cm 25 0ºC SN Units ppm mg L ppb µg L 1 234µS cm 25 0ºC SN Resolution 0 001 0 01 Instruction Manual Section 8 programming Measurements LIQ MAN 1066 April 2017 Programming Measurements 65 FIGuRE 8 5 Configuring Ozone Measurements ...

Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 77: ...ration instead of manual calibration Auto calibration avoids common pitfalls and reduces errors The transmitter recognizes the buffers and uses temperature corrected pH values in the calibration Once the 1066 successfully completes the calibration it calculates and displays the calibration slope and offset The slope is reported as the slope at 25 C To calibrate the pH loop with a connected pH sens...

Страница 78: ...Buffer Cal Menu The following screens may appear if the auto cal is unsuccessful 1 A High Slope Error will generate this screen display 2 A Low Slope Error will generate this screen display 3 An Offset Error will generate this screen display 9 2 1 1 234 µS cm 25 0 ºC SN pH Cal Buffer Cal Standardize Slope 59 16mV pH Offset 600 mV 1 234 µS cm 25 0 ºC SN Setup Stable Time 10 sec Stable Delta 0 02 pH...

Страница 79: ...d in a buffer solution the measured pH will differ from the buffer pH by an amount equivalent to the standardization offset This screen may appear if pH Cal is unsuccessful An Offset Error will generate this screen display If the pH Cal is successful the screen will return to the Cal sub menu SMART sensor auto calibration upload pH All calibration data including slope mV pH unit offset mV glass im...

Страница 80: ...following sub sections show the initial display screen that appears for each calibration routine Use the flow diagram for ORp calibration at the end of Section 8 and the live screen prompts to complete calibration Standardization ORP and Redox For process control it is often important to make the measured ORP and Redox agree with the ORP and Redox of a standard solution During calibration the meas...

Страница 81: ...istance of these solutions may be too low for an accurate measurement Calibrate sensors with 0 01 cm cell constant using method c c To calibrate a 0 01 cm sensor check it against a standard instrument and 0 01 cm sensor while both sensors are measuring water having a conductivity between 5 and 10 µS cm To avoid drift caused by absorption of atmospheric carbon dioxide saturate the sample with air b...

Страница 82: ...fter selecting Zero Cal from the Conductivity Calibration screen The adjacent screen will appear if zero Cal is successful The screen will return to the conductivity Cal Menu The adjacent screen may appear if zero Cal is unsuccessful Calibrating the Sensor in a Conductivity Standard in process cal This procedure is used to calibrate the sensor and transmitter against a solution of known conductivi...

Страница 83: ... sensor If the meter cal is successful the screen will return to the conductivity Cal Menu The adjacent screen will appear if Meter Cal is unsuccessful The screen will return to the conductivity Cal Menu Cal Factor Upon initial installation and power up if 4 electrode was selected for the sensor type in the Quick Start menus the user enters a Cell Constant and a Cal Factor using the instrument key...

Страница 84: ...librate toroidal conductivity 1 Press the MENU button 2 Select Calibrate Press ENTER 3 Select Conductivity Press ENTER The adjacent screen will appear To calibrate Toroidal Conductivity or Temperature scroll to the desired item and press ENTER The following sub sections show the initial display screen that appears for each calibration routine Use the flow diagram for Conductivity calibration at th...

Страница 85: ...ty reading of the 1066 to ensure that the reading is accurate This is done by submerging the probe in the sample of known conductivity then adjusting the displayed value if necessary to correspond to the conductivity value of the sample The probe must be cleaned before performing this procedure The temperature reading must also be checked and standardized if necessary prior to performing this proc...

Страница 86: ...d current before converting the result to a chlorine value New sensors require zeroing before being placed in service and sensors should be zeroed whenever the electrolyte solution is replaced To calibrate the chlorine sensor access the Calibration screen by pressing ENTER from the main screen and press ENTER The following calibration routines are covered 1 Zero Cal Zeroing the sensor in solution ...

Страница 87: ...ine In the second step the treated sample flows to the sensor The sensor is a membrane covered amperometric sensor whose output is proportional to the concentration of iodine Because the concentration of iodine is proportional to the concentration of total chlorine the transmitter can be calibrated to read total chlorine Because the sensor really measures iodine calibrating the sensor requires exp...

Страница 88: ...ion at the end of Sec 9 and the live screen prompts to complete calibration This adjacent screen appears after selecting Total Chlorine calibration Zeroing the Sensor The adjacent screen will appear during Zero Cal Be sure sensor has been running in zero solution for at least two hours before starting zero step The adjacent screen will appear if In Zero Cal is successful The screen will return to ...

Страница 89: ...bration curve Because stable monochloramine standards do not exist the sensor must be calibrated against a test run on a grab sample of the process liquid Several manufacturers offer portable test kits for this purpose To calibrate the chlorine sensor access the Calibration screen by pressing ENTER from the main screen and press ENTER The following calibration routines are covered 1 Zero Cal Zeroi...

Страница 90: ...ual current The transmitter compensates for the residual current by subtracting it from the measured current before converting the result to a dissolved oxygen value New sensors require zeroing before being placed in service and sensors should be zeroed whenever the electrolyte solution is replaced The recommended zero standard is 5 sodium sulfite in water although oxygen free nitrogen can also be...

Страница 91: ...ated air The transmitter monitors the sensor current When the current is stable the transmitter stores the current and measures the temperature using a temperature element inside the oxygen sensor The user must enter the barometric pressure From the temperature the transmitter calculates the saturation vapor pressure of water Next it calculates the pressure of dry air by subtracting the vapor pres...

Страница 92: ...pear if In Zero Cal is successful The screen will return to the Amperometric Cal Menu The adjacent screen may appear if In Zero Cal is unsuccessful The screen will return to the Amperometric Cal Menu Calibrating the Sensor in Air The adjacent screen will appear prior to Air Cal The adjacent screen will appear if In Air Cal is successful The screen will return to the Amperometric Cal Menu The adjac...

Страница 93: ...rom the measured current before converting the result to an ozone value New sensors require zeroing before being placed in service and sensors should be zeroed whenever the electrolyte solution is replaced The best zero standard is deionized water The purpose of the In Process Calibration is to establish the slope of the calibration curve Because stable ozone standards do not exist the sensor must...

Страница 94: ...ful The screen will return to the Amperometric Cal Menu In Process Calibration The following screen will appear after selecting In Process Cal If the In Process Cal is successful the screen will return to the Cal sub menu The following screen may appear if In Zero Cal is unsuccessful The screen will return to the Amperometric Cal Menu 9 8 1 9 8 2 1 234 nA SN Zero Cal Zeroing Wait 1 234 µS cm 25 0 ...

Страница 95: ... button 2 Select Calibrate Press ENTER 3 Select Temperature Press ENTER The adjacent screen will appear The following sub section provides you with the initial display screen that appears for temperature calibration Use the flow diagram for Temp calibration at the end of Sec 7 to complete calibration Calibration The adjacent screen will appear during Temperature Cal If the sensor Temperature offse...

Страница 96: ...ffer 1 Buffer 2 1 234µS cm 25 0ºC SN Manual Cal Cal in progress Please wait 1 234µS cm 25 0ºC SN pH Manual Cal Slope 59 16mV pH Offset 60 mV 1 234µS cm 25 0ºC SN pH Buffer Cal Auto Manual 1 234µS cm 25 0ºC SN pH Auto Cal Start AutoCal Setup 1 234µS cm 25 0ºC SN Setup Stable Time 10 sec Stable Delta 0 02 pH Buffer Standard 1 234µS cm 25 0ºC SN pH Auto Cal Place Sensor in Buffer 1 Press ENTER 1 234µ...

Страница 97: ...it 1 234µS cm 25 0ºC SN Zero Cal In Air In Water 1 234µS cm 25 0ºC SN InProcess Cal Press ENTER if reading is stable 1 234µS cm 25 0ºC SN InProcess Cal Cal in progress Please wait 1 234µS cm 25 0ºC SN InProcess Cal Updated cell constant 1 00135 cm 1 234µS cm 25 0ºC SN InProcess Cal Wait for stable reading 1 234µS cm 25 0ºC SN Zero Cal Sensor must be dry and in air 1 234µS cm 25 0ºC SN InProcess Ca...

Страница 98: ...Process Cal 1 234µS cm 25 0ºC SN Enter Value 10 00 ppm 1 234µS cm 25 0ºC SN Zero Cal Sensor zero done 1 234µS cm 25 0ºC SN Zero Cal Zeroing Wait 1 234µS cm 25 0ºC SN InProcess Cal Wait for stable reading 1 234µS cm 25 0ºC SN InProcess Cal Press ENTER if reading is stable 1 234µS cm 25 0ºC SN InProcess Cal Take sample Press ENTER Free Chlorine Total Chlorine Chloramine M A I N M E N U C a l i b r a...

Страница 99: ... zero done 1 234µS cm 25 0ºC SN InProcess Cal Cal in progress Please wait 1 234µS cm 25 0ºC SN Enter Value 10 00 ppm 1 234µS cm 25 0ºC SN InProcess Cal Wait for stable reading 1 234µS cm 25 0ºC SN InProcess Cal Press ENTER if reading is stable 1 234µS cm 25 0ºC SN InProcess Cal Take sample Press ENTER 1 234µS cm 25 0ºC SN Air Cal Start Calibration Setup 1 234µS cm 25 0ºC SN Setup Stable Time 10 se...

Страница 100: ...tal Chlorine Chloramine Ozone Oxygen pH ORP Redox Conductivity 1 234µS cm 25 0ºC SN Calibration Zero Cal In Process Cal 1 234µS cm 25 0ºC SN Enter Valu e 10 00 ppm 1 234µS cm 25 0ºC SN Zero Cal Zeroing Wait 1 234µS cm 25 0ºC SN Zero Cal Sensor zero done 1 234µS cm 25 0ºC SN InProcess Cal Wait for stable reading 1 234µS cm 25 0ºC SN InProcess Cal Press ENTER if reading is stable 1 234µS cm 25 0ºC S...

Страница 101: ...one Oxygen pH ORP Redox Conductivity 1 234µS cm 25 0ºC SN ISE Cal Standardize 2 Pt Calibration Slope 0mV decade Offset 0 mV Setup 1 234µS cm 25 0ºC SN Enter Value 0600 mV 1 234µS cm 25 0ºC SN Stable Time 010 sec 1 234µS cm 25 0ºC SN Stable Delta 0001 mV 1 234µS cm 25 0ºC SN Setup Stable Time 10sec Stable Delta 1 mV 1 234µS cm 25 0ºC SN Standard M Stabilizing Wait 1 234µS cm 25 0ºC SN 2 Point Cal S...

Страница 102: ...hlorine Chloramine Ozone Oxygen pH ORP Redox Conductivity 1 234µS cm 25 0ºC SN Calibrate Cal in progress Please wait 1 234µS cm 25 0ºC SN Temp Offset 5 C Continue No Yes 1 234µS cm 25 0ºC SN Calibrate 025 0 C Temperature M A I N M E N U C a l i b r a t e Section 9 Calibration Instruction Manual April 2017 LIQ MAN 1066 FIGuRE 9 7 Calibrate Temperature 92 Calibration ...

Страница 103: ...osen the Device Revision of the 1066 is Device Revision 1 when haRT 7 is chosen the Device Revision is Revision 2 or greater The Device Revision of the DD Device Description and install files for AMS and DeltaV used should be the same as the Device Revision of the 1066 A single HART 5 Device Revision 1 or HART 7 Device Revision 2 or greater DD Device Description is used for all model codes of the ...

Страница 104: ...ain Menu If HART 7 is chosen Univ Cmd Rev 7 the following controls are available univ Cmd Rev toggles between HART version 5 and HART version 7 If the HART host being used can accommodate HART 7 HART 7 should be chosen due to its larger feature set If the host can only use HART 5 then HART 5 must be chosen Note If the 1066 is connected to a HART host and the HART version is changed the host will l...

Страница 105: ...ose 0 unless Multidrop is being used If Multidrop is being used each transmitter should have its own polling address of from 1 to 15 Burst Mode Toggles the single HART 5 burst message on or off See below for the HART burst commands available Burst Commands available in haRT 5 and haRT 7 If burst messages are enabled by setting the burst messages to on the information in the burst message can be se...

Страница 106: ...ce variable other than the Primary variable is out of its limits Bit 2 loop Current Saturated This bit is set when Analog Output 1 is not fixed and it is less than 3 8 mA or greater than 22 0 mA Bit 3 loop Current Fixed This bit is set when Analog Output 1 is being simulated calibrated or when a device failure is detected and the Analog Output 1 is configured to output a fixed value Bit 4 More Sta...

Страница 107: ...on of events such as calibration in progress and relay activation are also indicated by status bits appendix 10 2 shows these bits organized according to the 1066 transmitter measurement type haRT hosts A HART host can access live measurements diagnostic messages and provide a tool for configuring the measurement and calibrating the 1066 The configuration parameters for the 1066 transmitter are li...

Страница 108: ...HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 FIGuRE 10 4 Device variables and Dynamic variables FIGuRE 10 5 Diagnostic Messages additional Transmitter Status ...

Страница 109: ...HART Communications 99 Instruction Manual Section 10 haRT Communication LIQ MAN 1066 April 2017 FIGuRE 10 6 Configuration FIGuRE 10 7 Calibration ...

Страница 110: ... Communication using the 1066 The 1066 can communicate by Wireless HART using the Smart Wireless THUM Adaptor and the 1420 Smart Wireless Gateway All the information available with the wired device can be accessed wirelessly making it possible to have the measurements and benefits of HART communication in locations where running cable would be difficult or prohibitively expensive Although HART 5 o...

Страница 111: ...ssignable to Dynamic variables variable Range primary value Type Conductivity 7 PV SV TV or QV 0 to 2000000 µS cm Resistivity 8 PV SV TV or QV 0 to 50000000 µ cm Concentration NaOH 9 PV SV TV or QV 0 to 12 HCl 10 PV SV TV or QV 0 to 15 Low H2SO4 11 PV SV TV or QV 0 to 25 High H2SO4 12 PV SV TV or QV 96 to 99 7 NaCl 13 PV SV TV or QV 0 to 25 Custom Concentration 14 PV SV TV or QV 0 to 1000 ppm 0 to...

Страница 112: ... current PV SV TV or QV 0 to 100000 nA Device variable Name assignable to Dynamic variables variable Range primary value Type Ozone 18 PV SV TV or QV 0 to 20 ppm 0 to 1000 ppb 0 to 20 mg L 0 to 1000 μg L Other Device variables Temperature PV SV TV or QV 15 to 200 C 5 to 360 F Temperature resistance TV or QV 0 to 100000 μ Sensor input current PV SV TV or QV 0 to 100000 nA Device variable Name assig...

Страница 113: ...erform its other functions normally 4 More Status available 3 1 user Data Error Warning An error was detected in the non volatile memory One or more user con figuration parameter may be corrupted Reset analyzer to factory defaults and re configure the device If the problem persists replace device 4 More Status available 3 2 Need Factory Calibration Warning The device s non volatile memory has been...

Страница 114: ...ake Output 2 out of hold mode 4 More Status available Byte Bit Meaning Class Device Status Bits Set 0 0 Temperature Error Error The temperature measuring circuit is open or shorted Check the wiring and the temperature element in the sensor If the temperature element in the sen sor is open or shorted replace the sensor 4 More Status available 7 Field Device Malfunction 2 0 Temperature high Warning ...

Страница 115: ...slope limit 1 Check the buffers used and retry buffer calibration 2 Increase the maximum slope limit default is 62 mV pH Note A slope of 62 mV pH or greater indicates that there has been an error made during calibration or a faulty pH sensor 3 Replace sensor 4 More Status available 4 1 ph Slope Too low Warning The pH slope calculated during buffer calibration was below the minimum slope limit The ...

Страница 116: ... curve is defined 4 More Status available 2 7 Input Out Of Range Warning The input is outside the device measurement range the range limits are 1066C 2 electrode conductivity Conductance 500 mS 1066CT 4 electrode conductivity Conductance 3000 mS or Vcond 0mV 1066T Conductance 1500 mS 1 Check sensor wiring 2 Replace sensor 4 More Status available 4 3 Calibration Error Warning An error occurred in t...

Страница 117: ...too high resulting in a negative reading The sensor zero limits are PV 0 5 if unit is ppm or mg L PV 50 if unit is ppb or ug L PV 2 0 for Saturation PV 2 0 for Concentration in Gas PV 20 ppm for Concentration in Gas PV 30 mmHg for Partial Pressure Perform sensor a zero calibration 4 More Status available 4 3 Calibration Error Warning There has been a calibration error Check the sensor 4 More Statu...

Страница 118: ...cks access to calibration from the key pad display of the 1066 000 disables UINT 2 R W 0 loi_main_display_upper Selects the measurement displayed on the upper por tion of the main display ENUM Assign PV Blank R W Assign PV loi_main_display_center Selects the measurement displayed on the center por tion of the main display ENUM Compensating pH free Cl Temperature PV Blank R W Temperature loi_main_d...

Страница 119: ... a successful calibration FLOAT 40 62 mV pH R W 40 mV pH max_ph_slope The maximum value of the slope allowed for a successful calibration FLOAT 40 62 mV pH R W 62 mV pH ph_stabilization_time The period of time used to determine stability of a pH sensor during automatic calibration FLOAT 0 99 seconds R W 10 seconds ph_stabilization_value The pH change used to determine stability of a pH sensor duri...

Страница 120: ...igh_fault_setpoint The high setpoint that triggers a glass pH electrode impedance alert FLOAT 0 0 9 999 0 Mohm R W 1 500 Mohm diagnostics_switch Turns the glass and reference electrode impedance measurements on and off ENUM Off On R W Off glassz_temperature_correction Turns temperature correction of the glass impedance measurement on and off ENUM Off On R W Off glassz_measurement_type Toggles betw...

Страница 121: ... zero_offset_in_air The zero offset determined by a zero calibration in air FLOAT R only 0 0 µS zero_offset_in_soln The zero offset determined by a zero calibration in a solution FLOAT R only 0 00 custom_curve_num_data_points The number of conductivity and concentration points to be used to calculate a custom concentration curve ENUM 2 3 4 5 R W 3 Custom Curve Concentration Data points custom_curv...

Страница 122: ...rature compensation used by the transmitter ENUM Linear slope Neutral salt Cation Raw None R W Linear slope temp_comp_slope Provides a slope value used in linear temperature compensation FLOAT 0 2 9 99 C R W 2 0 C reference_temperature The temperature that temperature compensation corrects to It is usually 25 C FLOAT R W 25 C Type Enumeration Range Read Write Default toroidal_sensor_model Selects ...

Страница 123: ...ee chlorine measurement is used FLOAT 0 00 14 00 pH R W 7 00 pH Type Enumeration Range Read Write Default chlorine_type The resolution of the displayed ozone measurement ENUM 0 001 0 01 R W 0 001 Type Enumeration Range Read Write Default polar_voltage The polarization voltage used by the transmitter automatically set by sensor selection FLOAT R W 0 temp_coeff The temperature coefficient used to co...

Страница 124: ...g inHg psi atm kPa mbar bar R W mmHg oxygen_process_pressure The process pressure used by the transmitter to calculate percent oxygen in gas or saturation FLOAT R W 760 0 mmHg oxygen_air_pressure The barometric pressure used by the transmitter during an air calibration FLOAT R W 760 0 mmHg oxygen_units The available units used for the various oxygen measure ment types ENUM R W Concentration in liq...

Страница 125: ...c Setup 2 2 Manual Setup 2 2 1 Device Information 2 2 1 1 Tag 2 2 1 2 Long Tag 2 2 1 3 Descriptor 2 2 1 4 Message 2 2 1 5 Date 2 2 1 6 Current Date 2 2 1 7 Current Time 2 2 1 8 Set Date and Time 2 2 2 Measurement 1066 ph ORp 2 2 2 1 PV is 2 2 2 2 Preamp Location 2 2 2 3 Soln Temp Correct n pH only 2 2 2 4 Temp Coefficient PV is pH and Soln Temp Correct n is Custom 2 2 2 5 Resolution pH only 2 2 2 ...

Страница 126: ...it 2 2 6 3 Sensor Config 2 2 6 3 1 Sensor Type Contacting Model 2 2 6 3 2 Sensor Model Toroidal Model 2 2 6 3 3 Range 2 2 6 3 4 Cell Constant 2 2 6 3 5 Cell Factor 4 Electrode 2 2 6 3 6 RTD Offset Cell Constant 0 02 2 2 6 3 7 RTD Slope Cell Constant 0 02 2 2 6 4 Setup Custom Curve Custom Curve Measurement 2 2 6 5 Temp Comp 2 2 6 5 1 Temp Comp Type 2 2 6 5 2 Temp Slope 2 2 6 5 3 Reference Temp 2 2 ...

Страница 127: ... 3 Trigger Level Unit 2 2 9 2 3 4 Classification 2 2 9 2 3 5 Update Rate 2 2 9 2 3 6 Default Update Rate 2 2 9 3 Burst Message 3 2 2 9 3 1 Burst Message 2 2 9 3 2 Message Content 2 2 9 3 3 Update Rate 2 2 9 3 3 1 Trigger Mode 2 2 9 3 3 2 Trigger Level 2 2 9 3 3 3 Trigger Level Unit 2 2 9 3 3 4 Classification 2 2 9 3 3 5 Update Rate 2 2 9 3 3 6 Default Update Rate 2 2 9 4 Event Notification 2 2 9 4...

Страница 128: ...ault Setpoint 2 3 1 2 Glass Impedance pH only 2 3 1 2 1 High Fault Setpoint 2 3 1 2 2 Temp Correction 2 3 1 2 3 Measurement Type 3 Service Tools 3 1 Alerts 3 2 Variables 3 3 Trends 3 4 Maintenance 3 4 1 1066 DO 1066 Dissolved Oxygen Only 3 4 1 1 Oxygen 3 4 1 2 Temperature 3 4 1 3 Input Current 3 4 1 4 Air Calibration 3 4 1 5 Air Calibration Setup 3 4 1 6 Zero Calibration 3 4 1 7 In Process Cal 3 4...

Страница 129: ...g Torridal Only 3 4 5 1 Conductivity 3 4 5 2 Raw Conductivity 3 4 5 3 Temperature 3 4 5 4 Zero Calibration 3 4 5 5 In Process Calibration 3 4 5 6 Cell Constant 3 4 5 7 Cell Factor 4 electrode 3 4 5 8 Zero Offset 3 4 5 9 Soln Offset Concentration 3 4 6 Temperature 3 4 6 1 Temperature 3 4 6 2 Calibrate Temp 3 4 6 3 RTD Offset 3 4 7 analog Outputs 3 4 7 1 Output 1 3 4 7 1 1 Loop Current 3 4 7 1 2 Cal...

Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...

Страница 131: ...ls in the same manner as they were received 4 Send the package prepaid to Emerson 8200 Market Blvd Chanhassen MN 55317 Attn Factory Repair RMA No ____________ Mark the package Returned for Repair Model No ___________ Non Warranty Repair The following is the procedure for returning for repair instruments that are no longer under warranty 1 Call Emerson for authorization 2 Supply the purchase order ...

Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...

Страница 133: ...Instruction Manual EC Declaration of Conformity LIQ MAN 1066 April 2017 EC Declaration of Conformity 123 Note Please see website for most recent Declaration ...

Страница 134: ...ompanies All other marks are the property of their respective owners The contents of this publication are presented for information purposes only and while effort has been made to ensure their accuracy they are not to be construed as warranties or guarantees express or implied regarding the products or services described herein or their use or applicability All sales are governed by our terms and ...

Отзывы: