Section 7: Transmitter programming
LIQ-MAN-1066
April 2017
7.4.4
7.5
7.5.1
7.5.2
Procedure: Ranging the Current Outputs
The adjacent screen will appear under Program/Output/Range.
Enter a value for 4 mA and 20 mA for each output. Follow the
to assign values to the outputs.
Setting a Security Code
Purpose
The security codes prevent accidental or unwanted changes to program settings, displays, and
calibration. The 1066 has two levels of security code to control access and use of the instrument
to different types of users. The two levels of security are:
•
all:
This is the Supervisory security level. It allows access to all menu functions, including
Programming, Calibration, Hold and Display.
•
Calibration/hold:
This is the operator or technician level menu. It allows access to only cali-
bration and Hold of the current outputs.
Procedure
1. Press MENU. The main menu screen appears. Choose
program
.
2. Scroll down to Security. Select Security.
3. The security entry screen appears. Enter a three digit security code for each of the desired
security levels. The security code takes effect two minutes after the last key stroke. Record the
security code(s) for future access and communication to operators or technicians as needed.
4. The display returns to the security menu screen. Press EXIT to return to the previous screen.
To return to the main display, press MENU followed by EXIT.
displays the security code screens.
1.234µS/cm 25.0 ºC
Output Range
O1 SN 4mA: 0.000 µS/cm
O1 SN 20mA: 20.00 µS/cm
O2 SN 4mA: 0 °C
O2 SN 20mA: 100 °C
1.234µS/cm 25.0ºC
Program
Outputs
Measurement
Temperature
Security
Diagnostic Setup
Reset Analyzer
Outputs
Program
Range
Configure
Simulate
1.234µS/cm 25.0ºC
Output Configure
Output1
Output2
1.234µS/cm 25.0ºC
Outputs
Range
Configure
Simulate
1.234µS/cm 25.0ºC
Output Range
O1 SN 4mA: 0.000
O1 SN 20mA: 20.00
O2 SN 4mA: 0ºC
S2 20mA: 100ºC
1.234µS/cm 25.0ºC
Simulate
Output 1
Output 2
1.234µS/cm 25.0ºC
Output Range
OM 4mA: 1.000
OM 20mA: 1.000%
OM 4mA: 1.000%
OM 20mA: 7.0 pH
1.234µS/cm 25.0ºC
OutputN Hold At
12.00mA
1.234µS/cm 25.0ºC
OutputM Configure
Assign: S1 Meas
Range: 4-20mA
Scale: Linear
Dampening: 0sec
MAIN MENU
1.234µS/cm 25.0ºC
OutputM Assign
S1 Measurement
S1 Temperature
1.234µS/cm 25.0ºC
OutputM Range
4-20mA
0-20mA
1.234µS/cm 25.0ºC
Scale
Linear
Log
FIGuRE 7-1. Configuring and
Ranging the Current Outputs
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...