Wiring
19
4.2.3
Current Output wiring
The 1066 HART units are shipped with two 4-20mA current outputs. Current Output 1 is loop
power; it is the HART communications channel. Current output 2 is available to report process
temperature measured by the temperature sensing element or RTD within the sensor.
Wiring locations for the outputs are on the main board which is mounted on the hinged door of
the instrument. Wire the output leads to the correct position on the main board using the lead
markings (+/positive, -/negative) on the board.
OUTPUT 2
TB5
TB3
TB2
TB4
TB1
+24V
GND
GND
+24V
T
H
U
M
ANODE
CATHODE
RTN
SNS
RTD IN
+V
-V
REF
SHLD
SENSOR WIRING
TB7
TB6
_
+
GND
SOL
SHLD
pH
(OUTPUT1)
LOOP PWR
4-
20
m
A
/ -2
4V
D
C
RE
TU
RN
4-2
0m
A
/ +
24
V
D
C
INSTALL PLUGS IN ALL OTHER
OPENINGS AS NEEDED
INNER ENCLOSURE
HINGE SIDE OF FRONT PANEL
1
TB7
/OUTPUT 2 REQUIRES EXTERNAL DC POWER.
2
TB6/
THUM TERMINAL IS USED ONLY FOR
WIRELESS THUM ADAPTOR INSTALLATIONS
.
1
2
1066 HART CIRCUIT BOARD
(pH/CL/DO/OZ)
ASSY 24406-xx
HINGED PANEL
DWG NO.
40106613
Instruction Manual
Section 4: Wiring
LIQ-MAN-1066
April 2017
FIGuRE 4-3. 1066 haRT loop power Wiring
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...