background image

6

PENBERTHY 

SERIES U FLAT GLASS GAGES 

INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

8 REMOVAL - DISASSEMBLY - REASSEMBLY

WARNING

Use only qualified, experienced personnel who 
are familiar with liquid level gage equipment and 
thoroughly understand the implications of the 
tables and all the instructions. DO NOT proceed 
with any maintenance unless the liquid level gage 
has been relieved of all pressure or vacuum, has 
been allowed to reach ambient temperature and 
has been drained or purged of all fluids. Failure 
to follow instructions can cause serious personal 
injury and property damage.

8.1 Disassembly

Secure a workbench longer than the liquid level 
gage, and sufficiently wide to lay out parts as 
they are removed.
1.  Hold gage firmly; loosen cap screws and 

nuts starting at both ends of each section 
and then proceed from both ends to the 
center of each section as shown in Figure 2.

 

Note:

 if maintenance is to be performed on 

one (1) side of the gage only, the cap screws 
on the opposite side of the gage do not need 
to be removed. If these screws are removed, 
then all glass, gaskets and cushions on 
both sides of the gage must be destroyed, 
disposed of and replaced.

2.  Nut loosening sequence
  -  Remove nuts, cap screws, washer and 

nameplate.

  -  Tap covers with rubber hammer as needed 

to loosen and remove.

  - Remove cushions, glass and gaskets.
  -  Tap liquid chamber or remaining covers as 

necessary with rubber hammer to break 
loose and remove remaining components.

  -  Remove, destroy and dispose of all 

glass, cushions and gaskets. Under no 
circumstances should these components 
be re-used or installed on a gage.

 

Note:

 if size of gage is smaller than shown, 

follow spiraling sequence from the ends 
until all bolting is loosened.

WARNING

Once used, cushions and gaskets are deformed 
permanently by compression and, if re-used, 
may cause leaks and high stress points resulting 
in glass breakage. Glass may contain hidden 
damage and internal stresses caused by previous 
usage. If re-used, the glass may break under 
pressure causing severe personal injury or 
property damage.

FIGURE 2
Nut loosening sequence

8.2 Inspection of glass seating surfaces

Clean the glass seating surfaces on the liquid 
chamber and cover with a soft metal scraper 
(preferably brass) to remove all burrs, rust and 
remnants of the previous gaskets and cushions. 
Exercise extreme care to avoid gouging or 
scarring gasket and cushion seating surfaces.
Use a known flat piece of metal the same 
approximate length as the glass or a new 
piece of glass and a thickness gage to check 
flatness of each glass seating surface on liquid 
chamber and under cover. Surfaces on the 
liquid chamber must be flat within 0.001 inch 
(0.03 mm) and on the cover must be flat within 
0.002 inch (0.05 mm). If any one surface is 
found to be beyond the tolerances described 
above, the entire gage must be disposed of and 
replaced.

WARNING

Flatness of glass seating surfaces outside the 
tolerances specified above is an indication of the 
gage having been overstressed through repeated 
exposure to mechanical, thermal or hydraulic 
shock during its previous service. Operation of a 
liquid level gage which has been overstressed will 
result in abnormal stresses on the glass which 
may cause glass to break with resulting sudden 
release of pressure, leakage of contained fluid, 
serious personal injury or property damage.

Glass seating surfaces should NOT be 
machined to achieve seating tolerance. The 
chamber and cover are designed for a critical 
thickness to achieve the pressure/temperature 
ratings. Machining glass seating surfaces may 
result in non-compliance to the necessary 
critical thickness due to material removal.

Содержание PENBERTHY Series

Страница 1: ...essed or implied shall be made by or on behalf of Emerson unless made in writing and signed by the company s general manager or director of engineering PENBERTHY SERIES U FLAT GLASS GAGES INSTALLATION OPERATION AND MAINTENANCE INSTRUCTIONS valves emerson com 2017 Emerson All rights reserved VCIOM 04042 EN 17 05 1 ABOUT THE MANUAL This manual has been prepared as an aid and guide for personnel invo...

Страница 2: ... 4 5 7 6000 psig 41370 kPa at 20 F 29 C to 250 F 121 C FKM Viton w 302 STS retainer 1 3 4 5 7 6000 psig 41370 kPa at 20 F 29 C to 400 F 204 C PTFE Teflon special chamber required 1 3 4 5 7 6000 psig 41370 kPa at 20 F 29 C to 500 F 260 C NOTE Lower temperatures are possible with metallic material variation e g 316 stainless construction PTFE Teflon gaskets cushions usable to 325 F 198 C To determin...

Страница 3: ... construction of the liquid level gage are compatible with both the contained media and surrounding atmosphere in the specific application 5 INSTALLATION Installation should only be undertaken by qualified experienced personnel who are familiar with equipment of this type They should have read and understood all of the instructions in this manual The user should refer to applicable technical data ...

Страница 4: ...pansion loops in the system Failure to allow for expansion or contraction can result in leaks or glass breakage 5 3 Mirror viewing For added safety a system of indirect viewing by means of mirrors should be installed to protect personnel from the hazards of possible gage failure 5 4 Nut retorquing Nut retorque is vital to the operation of a liquid level gage because gaskets take permanent set unde...

Страница 5: ...er and a soft cloth Inspect the surface of the glass for any clouding etching or scratching or physical damage such as bruises checks or corrosion Glass that is damaged is weakened and may break under pressure 7 2 Troubleshooting Problem Glass becomes etched prematurely or clouded in service Cause Fluid being handled is not compatible with the glass Solution Check application to determine actual o...

Страница 6: ...ING Once used cushions and gaskets are deformed permanently by compression and if re used may cause leaks and high stress points resulting in glass breakage Glass may contain hidden damage and internal stresses caused by previous usage If re used the glass may break under pressure causing severe personal injury or property damage FIGURE 2 Nut loosening sequence 8 2 Inspection of glass seating surf...

Страница 7: ... Install nameplate washer and nuts to studs on the opposite side of the gage Tighten nuts with fingers Using a torque wrench tighten cap screws and nuts in 5 ft lb 7 Nm increments following the sequence in Figure 1 until the torque values shown in Table 2 are reached Note depending on gage size there may be less bolting than shown in Figure 1 Start at the center and follow Z pattern outward to the...

Страница 8: ...plate 331 Band RECOMMENDED SPARE PARTS Item Description Min qty 3 Stud 2 per sect 4 Nut 2 per ect 48 Glass 1 100 Screw 1 per sect 7 Gasket 2 8 Cushion 2 FIGURE 3 U Transparent PENBERTHY SERIES U FLAT GLASS GAGES INSTALLATION OPERATION AND MAINTENANCE INSTRUCTIONS NOTE Size 7 shown actual gage may be shorter and require fewer bolting components ...

Отзывы: