Table 8-1. Diagnostic/Unit Alarm Fault Definitions – Membrane Keypad Only
LED
Flashes
Status
4-20 mA Line
Fault
Self-
Clearing
HEATER T/C
1
OPEN
Dependent on position 3 of SW2*
1
NO
2
SHORTED
Dependent on position 3 of SW2*
2
NO
3
REVERSED
Dependent on position 3 of SW2*
3
NO
4
A/D COMM ERROR
Dependent on position 3 of SW2*
4
NO
HEATER
1
OPEN
Dependent on position 3 of SW2*
5
NO
2
HIGH HIGH TEMP
Dependent on position 3 of SW2*
6
NO
3
HIGH CASE TEMP
Dependent on position 3 of SW2*
7
YES
4
LOW TEMP
Dependent on position 3 of SW2*
8
YES
5
HIGH TEMP
Dependent on position 3 of SW2*
9
YES
O2 CELL
1
HIGH mV
Dependent on position 3 of SW2*
10
YES
3
BAD
Track O
2
11
YES
4
EEPROM CORRUPT
Dependent on position 3 of SW2*
12
NO
CALIBRATION
1
INVALID SLOPE
Track O
2
13
YES
2
INVALID CONSTANT
Track O
2
14
YES
3
LAST CALIBRATION
FAILED
Track O
2
15
YES
**
CALIBRATION RECOM-
MENDED
Track O
2
YES
*Critical alarm conditions will render the O
2
measurement as unusable, and any of these events will cause the
4-20 mA signal to go to a user-selectable limit of 3.5 mA or 21.6 mA (position 3 of SW2). Factory default value
is 3.5 mA. Alarms which are not “self-clearing” will require recycling of power to the electronics.
**The CALIBRATION RECOMMENDED alarm flashes the Calibration Recommended alarm LED on the opera-
tor’s keypad.
Table 8-2. Diagnostic/Unit Alarm Fault Definitions – LOI
Message
Status
Fault
Number
Self-
Clearing
O2 T/C OPEN
HEATER T/C OPEN
1
NO
O2 T/C SHORTED
HEATER T/C SHORTED
2
NO
O2 T/C REVERSED
HEATER T/C POLARITY REVERSED
3
NO
ADC ERROR
A/D COMM ERROR
4
NO
O2 HEATER OPEN
O2 HEATER OPEN
5
NO
VERY HI O2 TEMP
VERY HIGH PROCESS TEMPERATURE
6
NO
BOARD TEMP HI
ELECTRONICS OVERHEATED
7
YES
O2 TEMP LOW
LOW PROCESS TEMPERATURE
8
YES
O2 TEMP HI
HIGH PROCESS TEMPERATURE
9
YES
O2 CELL OPEN
O2 CELL OPEN
10
YES
O2 CELL BAD
O2 CELL FAILED
11,13, 14
YES
EEPROM CORRUPT
EEPROM FAILED
12
NO
CALIB FAILED
LAST CALIBRATION FAILED
15
YES
LINE FREQ ERROR
INCORRECT INPUT LINE FREQUENCY
DETECTED ON POWER UP
NO