Operations Manual
FGD-MAN-0047
January 2017
21
7.4.2
Calibration failure
If the calibration procedure fails, the display shows Fail Cal, the Status
LED alternates red/green flashes and the analog output changes back
and forth from 3.0–3.3 mA.
The unit remains in a failed state until a
. After the manual
reset, the unit will return to normal operation based on previous
calibration values.