MRLDS-400 Series Application
Operation • 21
4.2.3.4. Reset to Factory Defaults
Instrument configuration may be reset to factory defaults via the smartphone
application:
• On the
Configure
tab select
Reset to Factory Default
and select
OK
to confirm.
• The instrument will automatically restart and disconnect from the smartphone
application.
WARNING! Resetting system to factory defaults will remove all custom system
configuration including unlock code and Bluetooth passcode. After system reset custom
unlock and Bluetooth passcodes should be configured to prevent unauthorized access and
reconfiguration of instrument.
4.2.3.5. Alarm Configuration
Low Alarm Setpoint
The value above which a low alarm condition occurs. The low alarm setpoint must
be less than the high alarm setpoint and greater than the low alarm limit. The low
alarm limit is the fixed minimum limit that is sensor–specific and not editable.
Range of acceptable setpoints is displayed when updating the parameter. To update
the setpoint:
• On the
Configure
tab select
Alarm
then
Low Alarm Setpoint
. Enter the new
setpoint and select
OK
to confirm.
IMPORTANT: In instruments with an oxygen sensor installed, low alarm behavior
operates in a depletion mode where gas measurements BELOW the low alarm setpoint
initiate a low alarm. This allows monitoring of oxygen displacement and enrichment
scenarios.
IMPORTANT: To prevent intermittent alarm operation at the setpoint due to measurement
noise this instrument implements hysteresis at the setpoint. Once the alarm level is
exceeded, the gas measurement must return a fixed percentage below the alarm threshold
before the alarm is disabled. Typical hysteresis value is set at 5% of full scale; however,
this is sensor-specific and non-editable.
High Alarm Setpoint
The value above which a high alarm condition occurs. The high alarm setpoint must
be less than the sensor full scale range and greater than the low alarm setpoint.
Range of acceptable setpoints is displayed when updating the parameter. To update
setpoint:
• On the
Configure
tab select
Alarm
then
High Alarm Setpoint
. Enter the new
setpoint and select
OK
to confirm.