14
Direct Connect Meters
Installation
3.3
Installing the core processor
for a diagram of the mounting bracket supplied with the core processor.
Both pipe mounting and wall mounting are shown.
Figure 3-1 Remote core processor – Wall mount or pipe mount
Table 3-3 Wire sizes and lengths – MVD Direct Connect installations with I.S. barrier
Span
Cable type
Wire size
Max length
Core processor
to barrier
RS-485
22 AWG (0,35 mm
2
) or larger 500 feet (150 meters)
Power
22 AWG (0,35 mm
2
)
300 feet (90 meters)
20 AWG (0,5 mm
2
)
500 feet (150 meters)
18 AWG (0,8 mm
2
)
500 feet (150 meters)
Barrier to host
RS-485
22 – 18 AWG
(0,35 – 0,8 mm
2
)
1000 ft (300 meters)
Barrier to power
supply
Power
22 AWG (0,35 mm
2
)
300 feet (90 meters)
20 AWG (0,5 mm
2
)
500 feet (150 meters)
18 AWG (0,8 mm
2
)
500 feet (150 meters)
Notes
• Core processor to barrier wire must be sized to provide a minimum of 15 V at the core processor.
See the discussion in the preceding section.
• Barrier to power supply wiring must be sized to provide a minimum of 19.2 V at the barrier. See
the discussion in the preceding section
Note
This step is required only if the core processor is mounted separately from the sensor. Refer to
and
Mounting bracket
(wall mount)
Mounting bracket
(pipe mount)
Содержание Micro Motion MVD Direct Connect
Страница 1: ...Installation Manual 20002273 Rev AA July 2014 Micro Motion MVD Direct Connect Meters ...
Страница 4: ...4 Direct Connect Meters ...
Страница 6: ...6 Direct Connect Meters Before you begin ...
Страница 26: ...26 Direct Connect Meters Installation ...
Страница 29: ...Installation Manual 29 MVD Direct Connect communications ...