24
Direct Connect Meters
Installation
3.8
Wiring to the remote host
1. At the remote host, open the wiring compartment and identify the RS-485 terminals.
Refer to the vendor documentation if required.
2. If you are connecting the RS-485 wires directly from the core processor (see Figures
2-1 and 2-2):
a. Connect the RS-485 wires from the core processor (see
) to the RS-485
terminals at the remote host.
b. Do not terminate the shield, braid, or drain wire(s) at the remote host.
c. Do not terminate the RS-485 lines using the standard 60-ohm termination resistor.
If possible, do not terminate the RS-485 lines at all. If the RS-485 cable is 1000 feet
(300 meters) long or longer, and termination is required, the total termination
must be 175 ohm or above.
3. If you are connecting the RS-485 wires from the I.S. barrier (see Figures 2-3 and 2-4):
a. Connect the RS-485 wires from the barrier (see
) to the RS-485 terminals
at the remote host.
b. Terminate the shields at the remote host.
c. The barrier contains internal pull-up/pull-down and termination resistors. Do not
add external resistors.
4. Close the wiring compartment.
3.9
Wiring to the power supply
1. You may connect multiple MVD Direct Connect installations to a single power supply,
as long as each installation receives sufficient power.
2. If you are connecting the power supply wires directly from the core processor (see
Figures 2-1 and 2-2):
a. Do not connect any other equipment to the power supply used for
MVD Direct Connect installations.
b. Connect the power supply wires from the core processor (see
matching positive and negative (+ and –).
If you are connecting the power supply wires from the I.S. barrier (see
):
a. The power supply may be used to power other equipment.
b. Connect the power supply wires from the barrier (see
), matching
positive and negative (+ and –).
Note
This step is required for all MVD Direct Connect installations.
Note
This step is required for all MVD Direct Connect installations.
Содержание Micro Motion MVD Direct Connect
Страница 1: ...Installation Manual 20002273 Rev AA July 2014 Micro Motion MVD Direct Connect Meters ...
Страница 4: ...4 Direct Connect Meters ...
Страница 6: ...6 Direct Connect Meters Before you begin ...
Страница 26: ...26 Direct Connect Meters Installation ...
Страница 29: ...Installation Manual 29 MVD Direct Connect communications ...