Alert data in transmitter memory
(continued)
Table 8-2:
Alert data structure
Transmitter action if condition occurs
Contents
Clearing
Recent Alerts
50 most recent alert postings or alert clearings Not cleared; maintained across transmitter
power cycles
8.6
Read totalizer and inventory values
ProLink III
View the desired variable on the main screen under
Process Variables
.
Overview
Totalizers keep track of the total amount of mass or volume measured by the transmitter
since the last totalizer reset. Inventories keep track of the total amount of mass or volume
measured by the transmitter since the last inventory reset.
Tip
You can use the inventories to keep a running total of mass or volume across multiple totalizer
resets.
8.7
Start and stop totalizers and inventories
ProLink III
Device Tools
>
Totalizer Control
>
Totalizer and Inventories
>
Start All Totals
Device Tools
>
Totalizer Control
>
Totalizer and Inventories
>
Stop All Totals
Overview
When you start a totalizer, it tracks process measurement. In a typical application, its value
increases with flow. When you stop a totalizer, it stops tracking process measurement and
its value does not change with flow. Inventories are started and stopped automatically,
when totalizers are started and stopped.
Important
Totalizers and inventories are started or stopped as a group. When you start any totalizer, all other
totalizers and all inventories are started simultaneously.When you stop any totalizer, all other
totalizers and all inventories are stopped simultaneously. You cannot start or stop inventories
directly.
Transmitter operation
Configuration and Use Manual
59
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...