![Emerson Micro Motion 5700 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/emerson/micro-motion-5700/micro-motion-5700_installation-manual_104641043.webp)
5
Wiring the power supply
You can install a user-supplied switch in the power supply line.
Important
For compliance with the Low Voltage Directive 2014/35/EU (European installations), a
switch in close proximity to the transmitter is required.
Procedure
1. Remove the wiring access cover.
2. Open the
Power
warning flap to locate the power terminals.
Figure 5-1: Location of power supply wiring terminals and equipment ground
A. Power supply wiring terminals (+ and -)
B. Equipment ground
3. Connect the power supply wires:
•
For DC power: connect to terminals
+
and
–
.
•
For AC power: connect to terminals
L/L1
(line) and
N/L2
(neutral).
4. Tighten the two screws holding the power connector in place.
5. Ground the power supply using the equipment ground, also under the
Power
warning flap.
Installation Manual
Wiring the power supply
MMI-20027478
February 2020
Installation Manual
43