![Emerson Micro Motion 5700 Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/emerson/micro-motion-5700/micro-motion-5700_installation-manual_104641027.webp)
4.4.2
Wire the mA/HART Output (externally powered)
Procedure
Wire to the appropriate output terminal and pins.
Figure 4-7: mA/HART Output wiring (externally powered)
A
1
2
D
C
B
A. mA/HART Output
B. 5–30 VDC (maximum)
for maximum loop resistance)
D. HART device
Figure 4-8: Externally-powered mA/HART Output: maximum loop resistance
0
100
200
300
400
500
600
700
800
900
1000
1100
0
7.5
15.0
22.5
30.0
B
A
A. Maximum resistance (Ω)
B. External supply voltage (V)
4.4.3
Wire the mA/HART multidrop installation (internally or
externally powered)
Procedure
See the following figure for information on wiring a mA/HART multidrop installation.
Installation Manual
Wiring the channels
MMI-20027478
February 2020
Installation Manual
27