![Emerson Micro Motion 4200 Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/emerson/micro-motion-4200/micro-motion-4200_installation-manual_104631017.webp)
•
Keep all ground leads as short as possible, less than 1 Ω impedance.
•
Connect ground leads directly to earth, or follow plant standards.
Procedure
1. Ground the sensor according to the instructions in the sensor documentation.
2. Ground the transmitter according to applicable local standards, using the
transmitter’s internal or external ground screw.
•
The earth ground terminal is located inside the power wiring compartment.
•
The external ground screw is located on the side of the transmitter located
below the transmitter tag.
2.5
Rotating the transmitter on the sensor
(optional)
For easier access to the user interface or the wiring terminals, the transmitter can be
rotated on the sensor in 45° increments, for eight different orientations.
Figure 2-8: Rotating the transmitter on the sensor
A. Clamping ring
Procedure
1. Remove the metal clamping ring from the base of the feed through (refer to
Installation Manual
Mounting and sensor wiring
MMI-20058013
April 2019
Installation Manual
17
Содержание Micro Motion 4200
Страница 1: ...Installation Manual MMI 20058013 Rev AB April 2019 Micro Motion 4200 2 Wire Transmitter ...
Страница 4: ...Contents Installation Manual April 2019 MMI 20058013 4 Micro Motion 4200 2 Wire Transmitter ...
Страница 10: ...Planning Installation Manual April 2019 MMI 20058013 10 Micro Motion 4200 2 Wire Transmitter ...
Страница 32: ...Wiring the channels Installation Manual April 2019 MMI 20058013 32 Micro Motion 4200 2 Wire Transmitter ...
Страница 34: ...Power up the transmitter Installation Manual April 2019 MMI 20058013 34 Micro Motion 4200 2 Wire Transmitter ...