C A - 3 0 0 0 , C A - 3 2 0 0
F
R O N T
P
A N E L
C
O N T R O L S
1.
Manual/Auto:
Changes the cycling mechanism to either manual or auto-
matic mode.
2.
Inhale
: Sets time interval (sec) for Inhale phase of automatic cycling.
Not operative in the manual mode.
3.
Exhale:
Sets time interval (sec) for Exhale phase of automatic cycling.
Not operative in the manual mode.
4.
Pause
: Sets time interval (sec) for Pause phase of automatic cycling.
Not operative in the manual mode.
5.
Power:
The power switch turns on or off the CoughAssist MI-E. The “
I
” symbol on the
switch designates the on position. When activated a green light within the switch will
illuminate. The “
O
” symbol designates the off position.
6.
Manual Control Lever:
Use to manually cycle the unit to inhale or exhale. Not operative
in the automatic mode.
7.
Pressure:
Varies the inhalation and exhalation pressures together (also see Inhale Pressure).
8.
Handle:
Recessed carrying handle.
9.
Patient Port:
Connection for patient circuit.
10.
: Symbol for Type B Equipment with F-Type applied part.
11.
Inhale Flow:
There are two inhalation flow settings: Full ( ) and Reduced ( ).
NOTE:
When using reduced inhalation flow there will be a small reduction in inhale pressure.
12.
Inhale Pressure:
Varies the inhalation pressure between 50% and 100% of the
exhale pressure.
13.
Pressure Gauge Zero Adjust:
Access to “zero” adjustment on the pressure gauge.
Use only if the gauge does not return to “O” when the unit is turned off (see Section 8).
14.
Pressure Gauge:
Indicates pressure in the patient circuit calibrated in cm H
2
O.
SECTION 4
7
C
O N T R O L S
,
C O N N E C T O R S
&
V I S U A L I N D I C AT O R S
o