Conductivity (uS) Calibration.
Conductivity calibration involves the following steps:
1) Working scale setup
Move cursor on “Range” then press wheel. Acording to probe reading capacity select proper scale by rotating wheel.
Once satisfied press wheel, move cursor on “OK” and press wheel again. See page 35 for proper scale settings. Also
available TDS conductivity with 0-9999 scale.
2) Turn off MAX5. Remove cover and locate CD module jumper settings as shown on page 35. Set jumpers as required.
Put cover back to MAX5.
3) Turn on MAX5. Once into conductivity menu choose Calib 1st Pt.Calibration.
Conductivity calibration procedure involves a zero calibration (1st point calib) and a 2nd calibration point that requires
a buffer solution with value near working range. During this procedure probe must be dry and clean and not installed in
plant. Once into “Calibration uS” menu move wheel on “Calib 1st Pt” then press wheel, move cursor on “OK” and press
wheel again.
Move wheel on “Calib 2nd Pt” then press wheel to enter into second point calibration submenu. Prepare buffer solution
and dip probe’s sensor on it. Wait until reading value is stable and according to buffer solution value move wheel until it
is the same on display (“uS default” field). End procedure moving cursor on “OK”
This procedure assumes that instrument is correctly configured. Otherwise unattended results may occurr.
fig. 38
fig. 39
fig. 40