background image

42

Содержание LDSPH PLUS Series

Страница 1: ...ains safety information that if ignored can endanger life or result in serious injury They are indicated by this icon Keep the instrument protected from sun and water Avoid water splashes Read Carefully ENGLISH Version R5 03 19 ERMES COMMUNICATION SYSTEM www ermes server com ...

Страница 2: ...d authorized personnel only Always discharge the liquid end before servicing the instrument Empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals Always read chemical safety datasheet Always wear protective clothing when handling hazardous or unknown chemicals Instrument must be operated serviced by trained technicians only All connection oper...

Страница 3: ...PH A PLUS Version stops probe s reading when into STANDBY mode and activates the cleaning probe procedure if enabled and for set time into SELF CLEAN MENU and restore time At the end of the STANDBY a new cleaning cycle and restore time re start 2 The wheel Located in the upper right side of LDSPH there is a wheel that must be used to control the instrument Wheel can be rotated in both directions t...

Страница 4: ...emove jumper resistor prior to install probe 11 10 Standby contact 11 12 pH Level 1 contact 19 18 pH Level 2 contact 14 Brown 15 Black 16 Blue 17 GND Proximity sensor mod SEPR dont remove jumper between blocks 16 and 17 21 GND 28 RS485 29 RS485 RS485 24 25 Opto coupled output pH PULSE To use with IS MF PLUS series dosing pumps 26 27 Opto coupled output pH PULSE 2 To use with IS MF PLUS series dosi...

Страница 5: ...nstrument activity For more information rotate the wheel when into main screen see next page During critical situations a warning alarm message may appear To in depth explanation completely rotate clockwise the wheel to review main instrument parameters and current outputs status sold as option Note the word PUMP as shown into this manual refers to a dosing device connected to the instrument WARNI...

Страница 6: ...atus Local Time Local Date pH probe reading Temperature probe reading Dosing alarm condition Probe failure status Alarm contact status Flow contact status Tank Level status Last pH calibration result Last pH calibration date Last Temp calibration result Last Temp calibration date Outputs Status See mainboard page 4 for related connetions ...

Страница 7: ...and enter the passcode Numbers can be selected rotating the wheel To set a new passcode choose PARAMETERS from Main Menu move on New Pcode click on wheel and enter a four numbers code Click on EXIT and choose YES to save request The new passcode is now ready Lost passcode Please dont forget the passcode if changed In the unfortunate event please call your local distributor for unlocking procedure ...

Страница 8: ...et see page 18 Dosing alarm see page 19 International see page 20 Probe Failure see page 21 Compensation see page 32 Flow see page 22 Service see page 22 Out of Range Alarm see page 24 mA Outputs see page 23 Communication see below Log Setup see page 38 Log View see page 38 Self Clean see page 25 Circulator Pump see page 26 Water Meter see page 29 Feed Forward see page 30 RS485 Setup see page 36 S...

Страница 9: ...N OFF mode while dosing ALKALI Set pH value at 7 00 OFF and 6 90 ON Set Pulse Speed per minute strokes per minute based on dosing device capabilities Instrument will leave the pH pump active until reading value will increase up to 7 00pH At 7 00pH the pH pump will be disabled until reading value will decrease under 6 90pH This mode is valid for all pH related outputs ON OFF mode while dosing ACID ...

Страница 10: ... For values of 7 5pH pump will be ON with 90 pulses per minute capacity The calculation is based on 180 pulses per minute setting see page 22 To end procedure move cursor on OK and press wheel to proceed to Save request screen Move wheel on YES to save or NO to discard changes Proportional Water Meter Proportional mode can also be set into PROP WM mode This option allows to regulate proportional i...

Страница 11: ...s within PWM operates For example set first pH value at 8 00 100 and second pH value at 4 0 0 For reading values to 8 00 the output will be permanently ON For reading values 4 0 the output will be permanently OFF For reading value of 7 00 the output will be OFF for 25 seconds ON for 75 seconds For reading value of 6 00 the output will be OFF for 50 seconds ON for 50 seconds Proportional Water Mete...

Страница 12: ...rmanently leave on for selected amount of time or off the output Parameters to set for this mode are pH range two pH values within PWM operates Ton ON period during output activity Toff OFF period during output activity For example set first pH value OFF at 7 20 Set second pH value ON at 8 00 Set working pause output activity withToff 0 seconds andTon 40 seconds For reading values to 8 00 the outp...

Страница 13: ...f a control valve a damper or the power supplied to a heating element To correctly setup PID as working mode configure the following parameters within two menus SETPOINT MODE PID and PID PARAMETERS 1 Choose related output to work into PID mode using Setpoint menu 2 Enter RANGE value Range is the maximum PID value over or under which depending on pH or pH the instrument will automatically switch in...

Страница 14: ...it Wait until reading value is stable and accor ding to buffer solution value move wheel until it is the same on display Cal at field Default value is 7 00pH To end procedure move cursor on OK and press wheel to proceed to next step Note buffer solution value may change if environment temperature it s different than 20 C Read solution s label for more information Accor ding to this occurrence pH D...

Страница 15: ...perature otherwise unattended results may occur Using an external thermometer read actual temperature and edit related field Cal at Confirm by pressing wheel To end procedure move cursor on OK and press whe el to proceed to Save request screen Move wheel on YES to save or NO to discard changes If an error oc curred during calibration procedure then the instrument will show an error message and wil...

Страница 16: ...elay for the pumps Delay occurs when instrument is powered Mode Move on Mode then press wheel If both pumps need to operate a startup priority can be set to allow the pH pump to begin to feed prior to Cl pump Choose pH priority to enable this function Cl pump will begin to dose when pH pump has stopped Tau If probes reading values are changing too fast increase TAU value to stabilize them Default ...

Страница 17: ...outputs Press wheel to move cursor on TIME field Once here choose a working time between 0 disabled or 199 minutes Move on EXIT then press wheel Choose YES to save changes Exit from main menu Main display will show a countdown for selected output To stop this countdown go back to Output Manager menu and choose AUTO as working mode or wait until countdown ends This function can be used for priming ...

Страница 18: ...et several parameters channel temperature instrument and setpoint Press wheel then change value to restore multiple choises are allowed to ON press wheel again move on OK then finally press wheel The instrument will go back to main menu and values will be restored ...

Страница 19: ...ead of a number minutes Dosing alarm can be set for both or just one pump E g To set Cl pump to stop after time ends and setpoint isn t still reached press wheel choose maximum time press wheel move on next field and choose STOP Time can be set between 0 and 100 minutes When satisfied with settings move on exit and press wheel To end procedure move cursor on OK and press wheel to proceed to Save r...

Страница 20: ...units format See table for differencies EUROPE IS Internationl Standard USA Date DD MMM YY Date MMM DD YY Time 24h Time AM PM C F Time Use this option to set local time Date Use this option to set date Move on exit to end changes To end procedure move cursor on OK and press wheel to proceed to Save request screen Move wheel on YES to save or NO to discard changes ...

Страница 21: ...on can be disabled selec ting OFF instead of a number minutes This function can be set for both or just one probe E g To set pH pump to stop after time ends and probe doesn t change reading values press wheel choose maxi mum time press wheel move on next field and choose STOP Time can be set between 100 and 254 minutes When satisfied with settings move on exit and press wheel To end procedure move...

Страница 22: ...pecified time when contact status changes To set it move wheel on Time 00 min click it and rotate to choose time from 0 to 99 minutes Confirm selection by clicking wheel To end procedure move cursor on OK and press wheel to proceed to Save request screen Move wheel on YES to save or NO to discard changes 17 Service This view only menu shows probes reading live and instrument ID for USB LOG connect...

Страница 23: ... disable output on alarm condition flow level probe failure dosage out of range Note The outputs available for the PID mA are those set in the menu SETPOINT PID outputs will be always disbled during an alarm condition Rotate wheel to move within all 3 channels Click wheel to selecte parameter and rotate wheel to change it Click wheel again and rotate wheel to move cursor on next parameter To end p...

Страница 24: ...cking on wheel and rotating it Press wheel again and move on next field Press wheel and enter a value for HIGH alarm Repeat procedure for pH Lo Dis and enter a value for LOW alarm As last option enter Time max 99 minutes after which if lower or higher read value condition stays then the alarm occurs to set into mode field To change alarm mode move wheel on Mode press it and choose between DOSE con...

Страница 25: ...een cleaning procedure and next procedure adjustable from 6 hours to 10 days Clean Time time required to complete probe cleaning procedure adjustable from 0 to 999 seconds Restore Time time to wait after cleaning to restore probe reading functionalities adjustable from 0 to 999 minutes Clean on Alarm Threshold alarm before to start procedure out of range alarm To find optimal values refers to prob...

Страница 26: ...e allows you to power a pump for water circulation inside the intake pipeline increasing the pressure To activate the circulation pump connected to terminals 5 E N see terminal set option to ENABLED To disable set the tool to DISABLED ...

Страница 27: ...er supply 85 264 VAC pH range 0 14 Environment Temperature 10 45 C 14 113 F Chemical Temperature 0 50 C 32 122 F Installation Class II Pollution Level 2 Packaging and Transporting Temperature 10 50 C 14 122 F Protection degree IP 65 ...

Страница 28: ...or configuration for two instruments Configuration of a Flow Switch with a voltage free contact and two instruments 14 15 16 17 SEPR brown blue black 14 15 16 17 N C contact OK N O contact Alarm brown blue black 14 15 16 17 14 15 16 17 NPED4 ...

Страница 29: ...will be notified into main screen and water meter status see page 6 FM Input can be set to operate Water Meter as 0 20mA or 4 20mA input P L pulses per liter or L P liters per pulse Tot WMI water passed through water meter FM Input Water Meter Working Mode RST TOT reset totalizer Timeout countdown to no water flow alarm 1 2 Connect as follows if mA working mode is enabled Block 1 red wire Block 2 ...

Страница 30: ...erms of time P depends on the present error I on the accumulation of past errors and D is a prediction of future errors based on current rate of change The weighted sum of these three actions is used to adjust the process via a control element such as the position of a control valve a damper or the power supplied to a heating element This function manages all instrument outputs based on changes re...

Страница 31: ...31 26 NOTES intentionally left blank page ...

Страница 32: ...ings varies from solution to solution and can be calculated using the following graphic Instrument has either fixed or adjustable automatic temperature compensation referenced to a standard temperature of 25 C To end procedure move cursor on OK and press wheel to proceed to Save request screen Move wheel on YES to save or NO to discard changes ...

Страница 33: ...33 mm 213 00 264 00 124 00 200 00 230 00 97 00 136 00 Appendix Dimensions ...

Страница 34: ...SM Module Located under mainboard cover there is a four pins connector that can be used to install USB ETHERNET or MODEM modules Modules come pre installed upon request and may appear different as shown different configurations SMS GMS module can be configured to send SMS messages containing critical instrument information To obtain reliable results with this feature please check the following lis...

Страница 35: ...g recording from LOG SETUP menu HOW TO RECORD INSTRUMENT S ACTIVITIES INTO USB PENDRIVE Insert USB pendrive into USB connector located on the right side of instrument Instrument will save data log on USB pendrive After succeded in saving data it will ask if delete instrument s log or not anyway USB pendrive will not be formatted Move wheel on YES to delete log info from instrument and return to ma...

Страница 36: ... alarm messages using its own modem sold as option It can be configured as follows SMS1 SMS2 SMS3 Using the wheel enter a mobile phone that will receive alert SMS messages if something wrong occurrs SMS number must be set using local number format For example 3391349134 will send an SMS message to mobile phone Log level and SMS frequency alert may be set using options in ACTIVE MSG within GSM menu...

Страница 37: ... lar to using a pay phone Unless there is a reason to receive a call the user does not care what number he or she is calling from There are times however when users who connect to the Internet using dynamic IP wish to allow other computers to locate them Perhaps they want to use CU SeeMe or use a VoIP application to make long distance phone calls using their IP connection In that case they would n...

Страница 38: ...ers For manual configuration option enter APN access point name and SIM phone number Move wheel on OK to save and move on ESC to go back to main menu Don t forget to enter SIM CODE into PIN NUMBER menu to unlock SIM WARNING THIS FUNCTION COULD NOT BE FREE OF CHARGE DEPENDING ON YOUR OPERATOR CONTRACT IT COULD GENERATE PAYING DATA TRAFFIC Email menu If Ethernet module or GPRS module is installed so...

Страница 39: ... INSTRUMENT WILL LOOSE DATE TIME Set ACTIVE to enabled to activate log recording TIME recording start time time format 23h e 59min EVERY recording frequency time format 23h e 59min Note advanced log control graph printing comparison tables event filtering etc is available through ERMES Communication Software for PC See ERMES Communication Software manual for proper PC software configuration LOG VI...

Страница 40: ...col and it is now a commonly available means of connecting industrial electronic devices From main menu select COMMUNICATION then MODBUS to access the options Set the communication speed according to the PLC system available Set the ID assigning an UNIQUE address to avoid conflicts 1 2 3 1 GND 2 A RS485 3 B RS485 To access the module MODBUS open the in strument only after power is switched off Nev...

Страница 41: ...t then click Otherwise click on SCAN to restart the search Enter WEP WPA WPA2 password if required and wait until connection has been estabilished and WiFi signal strength appears To obtain a reliable connection be sure to install the controller within WiFi range See your router features and installation procedure for best results WiFi signal strength ...

Страница 42: ...42 ...

Страница 43: ...l 11 8 4 PWM fixed Cl 12 8 5 PID Cl 13 9 Probe Calibration C 14 10 Parameters 16 11 Output Manager 17 12 Instrument Reset 18 13 Dosing Alarm 19 14 International 20 15 Probe Failure 21 16 Flow Contact 22 17 Service 22 18 mA Outputs 23 19 Out of range alarm 24 20 Self Clean 25 21 Circulator pump 26 22 Technical Information 27 23 SEPR configuration 28 24 Water Meter 29 25 Feed Forward and mA module 3...

Страница 44: ...rial types and send them according to local recycling disposal requirements We appreciate your efforts in supporting your local Recycle Environmental Program Working together we ll form an active union to assure the world s invaluable resources are conserved ...

Отзывы: