10. “Parameters”
From “Menu Calibration” choose “Parameters”. This menu allows to set a delay (max 60 minutes) before pumps begin
to feed. Furthermore use this menu to set pH pump startup priority and to change default passcode.
Feeding Delay.
Move on “Feeding Delay” then press wheel. Choose a value between 0 (disabled) and 60 minutes (maximum delay
time). This feature can be used to accord a startup delay for the pumps. Delay occurs when instrument is powered.
Mode.
Move on “Mode” then press wheel. If both pumps need to operate, a startup priority can be set to allow the pH pump to
begin to feed prior to Cl pump. Choose “pH priority” to enable this function. Cl pump will begin to dose when pH pump
has stopped.
Tau.
If probes reading values are changing too fast increase TAU value to stabilize them. Default value is 05. Maximum
value is 30.
New Pcode.
See page 10.
To end procedure move cursor on “OK” and press wheel to proceed to “Save” request screen. Move wheel on “YES” to
save or “NO” to discard changes.