background image

8

CALIBRATION:  2)CALIB

Choose “CALIB” from main menu. Display shows:

Calib

1)Zero

Press “UP” key. Display shows:

Calib

2) Slope

Zero calibration.

When display shows “1) Zero” press “E” to enter into zero calbration procedure. Display shows:

R 1.00Cl

C 0.00Cl

To calibrate it , if you are using an ECL4-6-7 probe, close water’s flow and wait for some minutes. If you are using an ECL1-
2-3-8 probe use flow-water without chlorine into probe holder or install an activated-carbon cartridge. If no one of these 
methods is vailable, unscrew top probe connector or, if not present, disconnect probe’s green wire (signal wire). Wait until “R” 

value is stable. 

ATTENTION: IT IS NOT NECESSARY  THAT “R”  VALUE WILL BE THE SAME OF “C” VALUE. 

Press “E” 

to confirm. Display shows:

Calibr.

Success.

Slope calibration.

When display shows “2) Slope” press “E” to enter into slope calbration procedure. Display shows:

R 1.00Cl

C 0.00Cl

To obtain reliable reading results is necessary to calibrate the probe using a chlorine  solution. Verify the plant’s chlorine value at  

the end of probe holder using a colorimetric (DPD1) analysis sistem. Using “UP” and “DOWN” keys enter this value in “C” field. 

Wait until “R” value is stable. 

ATTENTION: IT IS NOT NECESSARY  THAT “R”  VALUE WILL BE THE SAME OF “C” VALUE. 

SETUP

Содержание DIN Digital CL

Страница 1: ...ns important safety informations about installation and use of this equipment Ignoring this informations could result in injuries or damages It is strictly forbidden to use this equipment with radioactive chemicals OPERATING MANUAL FOR DIN DIGITAL CL CONTROLLER ...

Страница 2: ... authorized personnel only Always discharge the liquid end before servicing the instrument Empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals Always read chemical safety datasheet Always wear protective clothing when handling hazardous or unknown chemicals Instrument must be operated serviced by trained technicians only All connection opera...

Страница 3: ... VIEW ON MENU 6 SETTING SETPOINTS 1 SETPN 6 Out mA 7 CALIBRATION 2 CALIB 8 DELAY 3 Delay 9 RESERVED 4 Reserv 9 PASSWORD SETUP 1 New Pw 10 STANDBY SETUP 2 Stand 10 MANUAL WORKING MODE 3 Serv 11 ERROR MESSAGES 11 Warning proportional mode operates on 100 or 150 seconds basis ...

Страница 4: ...d data The Enter key confirms the selection Press RIGHT key for about 4 seconds to switch off the instrument the display shows OFF Repeat the sequence to return to the normal operating mode To reset the instrument restore default settings unplug power supply and keeping pressed UP and DOWN keys plug in power supply The instrument will show Checksum Error Press any key Press any key to continue Dis...

Страница 5: ... without a proximity sensor If needed proximity sensor turns the instrument into stand by mode if there isn t flow Instrument s display shows NO FLOW Connect SEPR to blocks 13 blue 15 black 16 brown Connect together block 14 with block 13 To work without a SEPR and with the same functionality use blocks 15 and 16 as a free contact Connect together block 14 with block 13 Using SEPR or the free cont...

Страница 6: ...l MAIN menu is configured as follows 1 Setup This menu is divided into 1 Setpn 2 Calib 3 Delay 2 Param This menu is divided into 1 New Pw 2 Stand 3 Serv service mode SETTING SETPOINTS 1 SETPN Out 1 and Out 2 are relay s driven outputs The instrument can be programmed to operate in On Off mode or Proportional mode Proportional mode can be selected in between 0 and 100 Select SETUP from MAIN menu Se...

Страница 7: ...N and OFF It is not possible to enter mixed values ex 100 for SP1 and ON for SP2 In this case it will not possible save them In proportional mode the instrument doesn t show values for setpoints status but only if they are on or off It is possible modify setpoint status on off and setpoint values with RIGHT key on the function to modify and using UP and DOWN key to change the value Enter the value...

Страница 8: ...ew top probe connector or if not present disconnect probe s green wire signal wire Wait until R value is stable ATTENTION IT IS NOT NECESSARY THAT R VALUE WILL BE THE SAME OF C VALUE Press E to confirm Display shows Calibr Success Slope calibration When display shows 2 Slope press E to enter into slope calbration procedure Display shows R 1 00Cl C 0 00Cl To obtain reliable reading results is neces...

Страница 9: ...eaches the setpoint values Default value is set to 0 Select DELAY from SETUP menu The instrument will show OUT1 10 Sec Press RIGHT key The instrument will show OUT2 5 Sec Delay time can be set from 0 no delay to 60 seconds Press E to confirm The display will show the confirmation message DELAYS SAVED ...

Страница 10: ...ss E to confirm The instrument will show the new password for 2 seconds and then will return to the main menu STANDBY SETUP 2 Stand For a good electrodes polarization a delayed startup every time the instrument is powered may be set Select 2 STAND from Param menu The display will show STANDBY 02 Sec It is possible to set the delay using UP and DOWN keys The time can be set between 0 no delay to 60...

Страница 11: ...ntact closed and N C contact open OFF is the status of N O contact open and N C contact closed ERROR MESSAGES If the display shows an error message use the following table ERROR 1 Buffer solution error during calibration Try again using a different buffer solution ERROR 2 Buffer solution read error during calibration Try again using a different buffer solution ERROR 3 N A ERROR 4 Offset error Veri...

Страница 12: ...ion center for the recycling of electronic and electrical equipment For more information visit the online site When dismantling a pump please separate material types and send them according to local recycling disposal requirements We appreciate your efforts in supporting your local Recycle Environmental Program Working together we ll form an active union to assure the world s invaluable resources ...

Отзывы: