Configuration
Full-duplex
Baud rate
1200…19200
Handshake
X-On / X-Off
Galvanic isolation
No
ESD stability
Min. 1000 V
Short-circuit protection
Yes
USB1 / USB2
USB host
USB full/low speed
Max. power load
400 mA (total, both connections)
PC
USB client
USB 1.1
Ethernet
Socket
RJ45
Speed
10/100 MBit/s
TTL-I/O
Socket
9-pin female D-sub
Inputs
2
Outputs
4
Galvanic isolation
No
CAN OUT
Socket
RJ12
Speed
500 kBit/s
Voltage
24 VDC ±10%
Current
Max. 500 mA
Stirrer / pump
Socket
5-pin mini-DIN
Pump detection
Yes
Stirrer detection
Yes
Pump voltage
24 V DC ±5 % (max. 400 mA)
Stirrer voltage range
0 to 18 V DC ±10 % (max. 300
mA)
SENSOR
Socket
LEMO triaxial, 9 mm
Current range
0 to 24 µA AC
Steps
0.1 µA
Limits of error
1 µA
Measuring range
± 2000 mV
Resolution
0.1 mV
Limits of error
2 mV
Burette drive
Drive
Stepping motor
Resolution
1/20000 of the burette volume
Limits of error
At 10% 15 µm
At 30% 15 µm
At 50% 25 µm
At 100% 50 µm
Filling time
Minimum 20 sec at 100% fill-
ing rate
Discharge time
Minimum 20 sec
Burette detection
Yes (RFID)
Volumetric Titrator
25
Technical data
Содержание Aquastar AQV21
Страница 1: ...Volumetric Titrator AQUASTAR AQV21 Installation Information...
Страница 2: ......
Страница 4: ......
Страница 23: ...Protective cover for touchscreen 51105567 Volumetric Titrator 23 Optional accessories...
Страница 31: ...10 Product life cycle model Volumetric Titrator 31 Product life cycle model...
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......